Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 95153

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
August 08, 2024
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
B. Braun Medical, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

AET36 ANESTH EXT SET - Anesthesia Extension Set - Normally Closed Dual Manifold, Four-way Stopcock, Model/Catalog Number: 472036

Z-3144-2024
Recall number
Z-3144-2024
Initiated
August 08, 2024
Classification
Class II
Status
Ongoing
Recalling firm
B. Braun Medical, Inc.
Quantity
19050 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Potential for product leakage from the Spin Lock Connector on the distal end of the device due to the migration of excess solvent during the assembly process.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process change control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Potential for product leakage from the Spin Lock Connector on the distal end of the device due to the migration of excess solvent during the assembly process.

Code information

Model No 472036; UDI-DI 04046964184659; Lots 0061898815, 0061905158, 00VL872690, 00VL882317 and 00VL882319

Distribution pattern

US Nationwide distribution.