Event 95171

7 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
August 15, 2024
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
B Braun Medical Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

7 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 7

SL BLOODLINE FOR FMC 2008 SERIES-Tubing to be used to transfer blood between a patient and a haemodialyser during haemodialysis. Model/Catalog Number: SL-2000M2095

Z-3173-2024
Recall number
Z-3173-2024
Initiated
August 15, 2024
Classification
Class II
Status
Ongoing
Recalling firm
B Braun Medical Inc
Quantity
10658351 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Incomplete insertion of the patient connector of the Streamline Airless System Hemodialysis Bloodline to the Locksite may result in the potential for occlusion during prograde rinseback procedures. If the patient connector is not fully inserted into the Locksite, the internal septum may not open fully enough to prevent restrictions to flow. may cause delays in therapy while clinicians are forced to switch to retrograde flow method

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Incomplete insertion of the patient connector of the Streamline Airless System Hemodialysis Bloodline to the Locksite may result in the potential for occlusion during prograde rinseback procedures. If the patient connector is not fully inserted into the Locksite, the internal septum may not open fully enough to prevent restrictions to flow. may cause delays in therapy while clinicians are forced to switch to retrograde flow method

Code information

UDI-DI 04046964367762; All Unexpired Lot Numbers

Distribution pattern

US Nationwide distribution.

device · product 2 of 7

STREAMLINE FRESENIUS FOR DAVITA-Tubing to be used to transfer blood between a patient and a haemodialyser during haemodialysis. Model/Catalog Number: SL-2000M2095D

Z-3174-2024
Recall number
Z-3174-2024
Initiated
August 15, 2024
Classification
Class II
Status
Ongoing
Recalling firm
B Braun Medical Inc
Quantity
4671540 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Incomplete insertion of the patient connector of the Streamline Airless System Hemodialysis Bloodline to the Locksite may result in the potential for occlusion during prograde rinseback procedures. If the patient connector is not fully inserted into the Locksite, the internal septum may not open fully enough to prevent restrictions to flow. may cause delays in therapy while clinicians are forced to switch to retrograde flow method

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Incomplete insertion of the patient connector of the Streamline Airless System Hemodialysis Bloodline to the Locksite may result in the potential for occlusion during prograde rinseback procedures. If the patient connector is not fully inserted into the Locksite, the internal septum may not open fully enough to prevent restrictions to flow. may cause delays in therapy while clinicians are forced to switch to retrograde flow method

Code information

UDI-DI 04046955348763; All Unexpired Lot Numbers

Distribution pattern

US Nationwide distribution.

device · product 3 of 7

STREAMLINE FRESENIUS FOR DAVITA-Tubing to be used to transfer blood between a patient and a haemodialyser during haemodialysis. Model/Catalog Number: SL-2000M2095DA

Z-3175-2024
Recall number
Z-3175-2024
Initiated
August 15, 2024
Classification
Class II
Status
Ongoing
Recalling firm
B Braun Medical Inc
Quantity
4984920 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Incomplete insertion of the patient connector of the Streamline Airless System Hemodialysis Bloodline to the Locksite may result in the potential for occlusion during prograde rinseback procedures. If the patient connector is not fully inserted into the Locksite, the internal septum may not open fully enough to prevent restrictions to flow. may cause delays in therapy while clinicians are forced to switch to retrograde flow method

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Incomplete insertion of the patient connector of the Streamline Airless System Hemodialysis Bloodline to the Locksite may result in the potential for occlusion during prograde rinseback procedures. If the patient connector is not fully inserted into the Locksite, the internal septum may not open fully enough to prevent restrictions to flow. may cause delays in therapy while clinicians are forced to switch to retrograde flow method

Code information

UDI-DI 04046955776023; All Unexpired Lot Numbers

Distribution pattern

US Nationwide distribution.

device · product 4 of 7

STREAMLINE BLOODLINE LONG VERSION, FMC-Tubing to be used to transfer blood between a patient and a haemodialyser during haemodialysis. Model/Catalog Number: SL-2000M2095L

Z-3176-2024
Recall number
Z-3176-2024
Initiated
August 15, 2024
Classification
Class II
Status
Ongoing
Recalling firm
B Braun Medical Inc
Quantity
2021376 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Incomplete insertion of the patient connector of the Streamline Airless System Hemodialysis Bloodline to the Locksite may result in the potential for occlusion during prograde rinseback procedures. If the patient connector is not fully inserted into the Locksite, the internal septum may not open fully enough to prevent restrictions to flow. may cause delays in therapy while clinicians are forced to switch to retrograde flow method

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Incomplete insertion of the patient connector of the Streamline Airless System Hemodialysis Bloodline to the Locksite may result in the potential for occlusion during prograde rinseback procedures. If the patient connector is not fully inserted into the Locksite, the internal septum may not open fully enough to prevent restrictions to flow. may cause delays in therapy while clinicians are forced to switch to retrograde flow method

Code information

UDI-DI 04046955348909; All Unexpired Lot Numbers

Distribution pattern

US Nationwide distribution.

device · product 5 of 7

STREAMLINE LONG-Tubing to be used to transfer blood between a patient and a haemodialyser during haemodialysis. Model/Catalog Number: SL-2000M2095W

Z-3177-2024
Recall number
Z-3177-2024
Initiated
August 15, 2024
Classification
Class II
Status
Ongoing
Recalling firm
B Braun Medical Inc
Quantity
142336 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Incomplete insertion of the patient connector of the Streamline Airless System Hemodialysis Bloodline to the Locksite may result in the potential for occlusion during prograde rinseback procedures. If the patient connector is not fully inserted into the Locksite, the internal septum may not open fully enough to prevent restrictions to flow. may cause delays in therapy while clinicians are forced to switch to retrograde flow method

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Incomplete insertion of the patient connector of the Streamline Airless System Hemodialysis Bloodline to the Locksite may result in the potential for occlusion during prograde rinseback procedures. If the patient connector is not fully inserted into the Locksite, the internal septum may not open fully enough to prevent restrictions to flow. may cause delays in therapy while clinicians are forced to switch to retrograde flow method

Code information

UDI-DI 04046955348800; All Unexpired Lot Numbers

Distribution pattern

US Nationwide distribution.

device · product 6 of 7

STREAMLINE BLOODLINE SET FOR DIALOG-Tubing to be used to transfer blood between a patient and a haemodialyser during haemodialysis. Model/Catalog Number: SL-2010M2096

Z-3178-2024
Recall number
Z-3178-2024
Initiated
August 15, 2024
Classification
Class II
Status
Ongoing
Recalling firm
B Braun Medical Inc
Quantity
9864165 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Incomplete insertion of the patient connector of the Streamline Airless System Hemodialysis Bloodline to the Locksite may result in the potential for occlusion during prograde rinseback procedures. If the patient connector is not fully inserted into the Locksite, the internal septum may not open fully enough to prevent restrictions to flow. may cause delays in therapy while clinicians are forced to switch to retrograde flow method

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Incomplete insertion of the patient connector of the Streamline Airless System Hemodialysis Bloodline to the Locksite may result in the potential for occlusion during prograde rinseback procedures. If the patient connector is not fully inserted into the Locksite, the internal septum may not open fully enough to prevent restrictions to flow. may cause delays in therapy while clinicians are forced to switch to retrograde flow method

Code information

UDI-DI 04046964367786; All Unexpired Lot Numbers

Distribution pattern

US Nationwide distribution.

device · product 7 of 7

STREAMLINE BLOODLINE SET FOR DIALOG DR-Tubing to be used to transfer blood between a patient and a haemodialyser during haemodialysis. Model/Catalog Number: SL-2010M2096A

Z-3179-2024
Recall number
Z-3179-2024
Initiated
August 15, 2024
Classification
Class II
Status
Ongoing
Recalling firm
B Braun Medical Inc
Quantity
11262112 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Incomplete insertion of the patient connector of the Streamline Airless System Hemodialysis Bloodline to the Locksite may result in the potential for occlusion during prograde rinseback procedures. If the patient connector is not fully inserted into the Locksite, the internal septum may not open fully enough to prevent restrictions to flow. may cause delays in therapy while clinicians are forced to switch to retrograde flow method

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Incomplete insertion of the patient connector of the Streamline Airless System Hemodialysis Bloodline to the Locksite may result in the potential for occlusion during prograde rinseback procedures. If the patient connector is not fully inserted into the Locksite, the internal septum may not open fully enough to prevent restrictions to flow. may cause delays in therapy while clinicians are forced to switch to retrograde flow method

Code information

UDI-DI 04046955674992; All Unexpired Lot Numbers

Distribution pattern

US Nationwide distribution.