Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 95215

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
August 12, 2024
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Becton, Dickinson and Company, BD Bio Sciences

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

BD FACSDuet Premium Sample Preparation System, REF: 666340, 666339; and BD FACSDuet Sample Preparation System (Base), REF: 663128, 662588, used with software V 1.4.1. Intended use: to prepare human specimens to be analyzed.

Z-0009-2025
Recall number
Z-0009-2025
Initiated
August 12, 2024
Classification
Class II
Status
Ongoing
Quantity
130

App-derived interpretation

Device software or design reason.device_software_design · v1.0.0
software may cause

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software change control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Sample preparation system with software may cause: 1) Multidispense feature enabled with 2-tube assay: Specimen volume dispensed into tube1 at least 3% higher than volume in tube2, or 2) Absolute counts calculated using Trucount Tubes: increased volume in tube1 may result in 20% higher absolute counts in tube 1 than tube 2; could result in additional blood collection, delayed results/patient therapy.

Code information

REF/UDI-DI/Serial Numbers: 666340/382906663405/9230010101, 9230010102, 9230010103, 9230010104, 9230010105, 9230010107, 9230010108, 9230010109, 9230010112, 9230010113, 9230010114, 9230010115, 9230010116, 9230010118, 9230010119, 9230010120. 666339/9230000102, 9230000103, 9230000104, 9230000105, 9230000106, 9230000112, 9230000113, 9230000114, 9230000118, 9230000120, 9230000121, 9230000123, 9230000124, 9230000126, 9230000128, 9230000130, 9230000132, 9230000134, 9230000135, 9230000138, 9230000144, 9230000145, 9230000131, 9230000152, 9230000149, 9230000030, 9230000101, 9230000108, 9230000109, 9230000110, 9230000111, 9230000115, 9230000116, 9230000117, 9230000119, 9230000122, 9230000125, 9230000127, 9230000129, 9230000136, 9230000139, 9230000141, 9230000142, 9230000143. 663128/382906631282/9220030004, 9220030005, 9220030008, 9220030016, 9220030017, 9220030027, 9220030029, 9220030031, 9220030043, 9220030053, 9220030057. 662588/9220010119, 9220010179, 9220010182, 9220010232, 9220010236, 9220010250, 9220010265, 9220010275, 9220010293, 9220010311, 9220010313, 9220010316, 9220010324, 9220010327, 9220010335, 9220010338, 9220010341, 9220010344, 9220010348, 9220010349, 9220010292, 9220010112, 9220010120, 9220010125, 9220010127, 9220010128, 9220010130, 9220010133, 9220010138, 9220010154, 9220010161, 9220010183, 9220010188, 9220010197, 9220010215, 9220010225, 9220010231, 9220010238, 9220010240, 9220010242, 9220010244, 9220010256, 9220010257, 9220010260, 9220010263, 9220010268, 9220010269, 9220010279, 9220010280, 9220010289, 9220010290, 9220010294, 9220010296, 9220010329, 9220010332, 9220010345, 9220010354, 9230000031, 99220010292.

Distribution pattern

US: NJ, UT, WI, TX, ID, AK, MN, MA, MD, TN, MI, CA, VA OUS: Ireland, Austria, Italy, Germany, Belgium, Finland, Canada, France, Spain, New Zealand, United Kingdom, Japan, Australia, Norway, Singapore, Sweden, Turkey, Brazil, Colombia, Netherlands, United Arab Emirates, Saudi Arabia, Mexico, Switzerland, Costa Rica, China, Azerbaijan, Algeria, Jordan, India