Exact FDA event identifier

Event 95254

3 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
August 15, 2024
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Siemens Medical Solutions USA, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

3 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 3

ACUSON Juniper 1.0, 1.5, 2.0, 2.5 Diagnostic Ultrasound Systems, REF: 11335791, with software. ACUSON Juniper Select 2.0, 2.5 Diagnostic Ultrasound Systems, REF: 11653093, with software.

Z-3215-2024
Recall number
Z-3215-2024
Initiated
August 15, 2024
Classification
Class II
Status
Ongoing
Quantity
7022

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
If ultrasound systems with software, are changed from factory default to : 1) Milliliters per second (ml/sec, mL/sec) or 2) Milliliters per minute (ml/min, mL/min); then systems will perform incorrect unit conversion of liters to milliliters using multiplier 100 instead of 1000, resulting values 10x smaller than actual, which could contribute misdiagnosis or negatively influence patient management

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

If ultrasound systems with software, are changed from factory default to : 1) Milliliters per second (ml/sec, mL/sec) or 2) Milliliters per minute (ml/min, mL/min); then systems will perform incorrect unit conversion of liters to milliliters using multiplier 100 instead of 1000, resulting values 10x smaller than actual, which could contribute misdiagnosis or negatively influence patient management

Code information

11335791, UDI-DI: 04056869152059: 1.0/VA10G and prior, 1.5/ VB10H and prior , 2.0/ VB11C and prior, 2.5/ VB30D and prior. 11653093, UDI-DI: 04056869957357: 2.0/ VB11C and prior, 2.5/ VB30D and prior.

Distribution pattern

US: MS, WA, TX, NJ, GA, PN, NY, AL, MN, NE, IN, CT, CA, IL, TN, WI, ND, NM, AR, NC, PR, RI, WY, MD, FL, HI, AZ, SC, MO, OK, VA, ME, KS, MI, LA, IA, OH, SD, NH, KY, AL, OR, MA, GU. OUS: Afghanistan, Albania, Algeria, Andorra, Angola, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belgium, Bolivia, Bosnia-Herzegovina, Botswana, Brazil, Brunei, Bulgaria, Cameroon, Canada, Chile, China, Colombia, Congo, the Democratic Republic of the, Costa Rica, C¿te dIvoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Eritrea, Ethiopia, Faroe Islands, Fiji, Finland, France, French Guinea, French Polynesia, Georgia, Germany, Greece, Guatemala, Guernsey, Guyana, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Islamic Republic of, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kosovo, Kuwait, Latvia, Lebanon, Libya, Liechtenstein LI B, Lithuania, Luxembourg, Macedonia, Malawi, Malaysia, Malta, Mexico, Moldova, Rep. of, Mongolia, Montenegro, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Republic Korea, Reunion, Romania, Ruanda, Russian Federation, San Marino, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Swaziland, Sweden, Switzerland, Taiwan, Tajikistan, Tanzania, Thailand, Trinidad,Tobago, Tunisia, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Vietnam, Yemen, Zambia, Zimbabwe

device · product 2 of 3

ACUSON Redwood 1.0 and 2.0 Diagnostic Ultrasound Systems, REF: 11503314, with software

Z-3216-2024
Recall number
Z-3216-2024
Initiated
August 15, 2024
Classification
Class II
Status
Ongoing
Quantity
1990

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
If ultrasound systems with software, are changed from factory default to : 1) Milliliters per second (ml/sec, mL/sec) or 2) Milliliters per minute (ml/min, mL/min); then systems will perform incorrect unit conversion of liters to milliliters using multiplier 100 instead of 1000, resulting values 10x smaller than actual, which could contribute misdiagnosis or negatively influence patient management

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

If ultrasound systems with software, are changed from factory default to : 1) Milliliters per second (ml/sec, mL/sec) or 2) Milliliters per minute (ml/min, mL/min); then systems will perform incorrect unit conversion of liters to milliliters using multiplier 100 instead of 1000, resulting values 10x smaller than actual, which could contribute misdiagnosis or negatively influence patient management

Code information

11503314, UDI-DI: 04056869251264: 1.0/Software VA10G and prior, 2.0/Software VA20K and prior

Distribution pattern

US: MS, WA, TX, NJ, GA, PN, NY, AL, MN, NE, IN, CT, CA, IL, TN, WI, ND, NM, AR, NC, PR, RI, WY, MD, FL, HI, AZ, SC, MO, OK, VA, ME, KS, MI, LA, IA, OH, SD, NH, KY, AL, OR, MA, GU. OUS: Afghanistan, Albania, Algeria, Andorra, Angola, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belgium, Bolivia, Bosnia-Herzegovina, Botswana, Brazil, Brunei, Bulgaria, Cameroon, Canada, Chile, China, Colombia, Congo, the Democratic Republic of the, Costa Rica, C¿te dIvoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Eritrea, Ethiopia, Faroe Islands, Fiji, Finland, France, French Guinea, French Polynesia, Georgia, Germany, Greece, Guatemala, Guernsey, Guyana, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Islamic Republic of, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kosovo, Kuwait, Latvia, Lebanon, Libya, Liechtenstein LI B, Lithuania, Luxembourg, Macedonia, Malawi, Malaysia, Malta, Mexico, Moldova, Rep. of, Mongolia, Montenegro, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Republic Korea, Reunion, Romania, Ruanda, Russian Federation, San Marino, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Swaziland, Sweden, Switzerland, Taiwan, Tajikistan, Tanzania, Thailand, Trinidad,Tobago, Tunisia, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Vietnam, Yemen, Zambia, Zimbabwe

device · product 3 of 3

ACUSON Maple 1.0 Diagnostic Ultrasound System, REF: 11711750, with software. ACUSON Maple Select 1.0 Diagnostic Ultrasound System, REF: 11711779, with software

Z-3217-2024
Recall number
Z-3217-2024
Initiated
August 15, 2024
Classification
Class II
Status
Ongoing
Quantity
321

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
If ultrasound systems with software, are changed from factory default to : 1) Milliliters per second (ml/sec, mL/sec) or 2) Milliliters per minute (ml/min, mL/min); then systems will perform incorrect unit conversion of liters to milliliters using multiplier 100 instead of 1000, resulting values 10x smaller than actual, which could contribute misdiagnosis or negatively influence patient management

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

If ultrasound systems with software, are changed from factory default to : 1) Milliliters per second (ml/sec, mL/sec) or 2) Milliliters per minute (ml/min, mL/min); then systems will perform incorrect unit conversion of liters to milliliters using multiplier 100 instead of 1000, resulting values 10x smaller than actual, which could contribute misdiagnosis or negatively influence patient management

Code information

11711750, UDI-DI: 04056869995809, Software VA10A 11711779, UDI-DI: 04056869995830, Software VA10A

Distribution pattern

US: MS, WA, TX, NJ, GA, PN, NY, AL, MN, NE, IN, CT, CA, IL, TN, WI, ND, NM, AR, NC, PR, RI, WY, MD, FL, HI, AZ, SC, MO, OK, VA, ME, KS, MI, LA, IA, OH, SD, NH, KY, AL, OR, MA, GU. OUS: Afghanistan, Albania, Algeria, Andorra, Angola, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belgium, Bolivia, Bosnia-Herzegovina, Botswana, Brazil, Brunei, Bulgaria, Cameroon, Canada, Chile, China, Colombia, Congo, the Democratic Republic of the, Costa Rica, C¿te dIvoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Eritrea, Ethiopia, Faroe Islands, Fiji, Finland, France, French Guinea, French Polynesia, Georgia, Germany, Greece, Guatemala, Guernsey, Guyana, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Islamic Republic of, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kosovo, Kuwait, Latvia, Lebanon, Libya, Liechtenstein LI B, Lithuania, Luxembourg, Macedonia, Malawi, Malaysia, Malta, Mexico, Moldova, Rep. of, Mongolia, Montenegro, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Republic Korea, Reunion, Romania, Ruanda, Russian Federation, San Marino, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Swaziland, Sweden, Switzerland, Taiwan, Tajikistan, Tanzania, Thailand, Trinidad,Tobago, Tunisia, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Vietnam, Yemen, Zambia, Zimbabwe