Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 95287

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
July 16, 2024
Product types
Device
Classifications
Class I
Statuses
Ongoing
Recalling firm wording
Smiths Medical ASD Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

CADD-Solis Li-ion Rechargeable Battery Packs, REF 21-2160-XX, which are also sold as a component of the following CADD Solis Ambulatory Infusion Systems: 21-2101-0100-50, 21-2101-0200-02, 21-2101-0200-03, 21-2101-0200-06, 21-2101-0200-07, 21-2101-0200-12, 21-2101-0200-14, 21-2101-0200-17, 21-2101-0200-238, 21-2101-249, 21-2101-51, 21-2102-0200-02, 21-2102-249, 21-2102-51, 21-2111-0100-00, 21-2111-0100-50, 21-2111-0100-51, 21-2111-0200-02, 21-2111-0200-03, 21-2111-0200-06, 21-2111-0200-07, 21-2111-0200-08, 21-2111-0200-12, 21-2111-0200-14, 21-2111-0200-17, 21-2111-0200-232, 21-2111-0200-50, 21-2111-0300-00, 21-2111-0300-01, 21-2111-0300-02, 21-2111-0300-03, 21-2111-0300-06, 21-2111-0300-07, 21-2111-0300-08, 21-2111-0300-09, 21-2111-0300-12, 21-2111-0300-14, 21-2111-0300-17, 21-2111-0300-231, 21-2111-0300-232, 21-2111-0300-50, 21-2111-0400-01, 21-2111-0400-51, 21-2111-0401-02L, 21-2111-0401-03L, 21-2111-0401-06L, 21-2111-0401-07, 21-2111-0401-07L, 21-2111-0401-08L, 21-2111-0401-12L, 21-2111-0401-14L, 21-2111-0401-17L, 21-2111-0401-249, 21-2111-0401-51, 21-2111-0401-51L, 21-2111-0401-78, 21-2111-0402-02, 21-2111-0402-02L, 21-2111-0402-03L, 21-2111-0402-06L, 21-2111-0402-07L, 21-2111-0402-08, 21-2111-0402-08L, 21-2111-0402-09_LOAN, 21-2111-0402-09-FUK, 21-2111-0402-09JP, 21-2111-0402-09-KYO, 21-2111-0402-09-NAG, 21-2111-0402-09-OKA, 21-2111-0402-09-OSA, 21-2111-0402-09-SAP, 21-2111-0402-09-SEN, 21-2111-0402-09-TOKE, 21-2111-0402-09-TOKW, 21-2111-0402-09-YOK, 21-2111-0402-12L, 21-2111-0402-14L, 21-2111-0402-17L, 21-2111-0402-249, 21-2111-0402-51, 21-2111-0402-51L, 21-2111-0402-78, 21-2112-0100-50, 21-2112-0100-51, 21-2112-0200-02, 21-2112-0200-06, 21-2112-0300-00, 21-2112-0300-01, 21-2112-0300-02, 21-2112-0300-03, 21-2112-0300-06, 21-2112-0300-231, 21-2112-0300-50, 21-2112-0400-01, 21-2112-0400-51, 21-2112-0401-02L, 21-2112-0401-03L, 21-2112-0401-07L, 21-2112-0401-08L, 21-2112-0401-17L, 21-2112-0401-51, 21-2112-0401-51L, 21-2112-0401-78, 21-2112-0402-02, 21-2112-0402-02L, 21-2112-0402-03L, 21-2112-0402-06L, 21-2112-0402-07L, 21-2112-0402-08L, 21-2112-0402-09JP, 21-2112-0402-09-SAI, 21-2112-0402-09-SEN, 21-2112-0402-09-TOKE, 21-2112-0402-17L, 21-2112-0402-249, 21-2112-0402-51, 21-2112-0402-51L, 21-2120-0100-02, 21-2120-0100-03, 21-2120-0100-06, 21-2120-0100-232, 21-2120-0100-233, 21-2120-0100-50, 21-2120-0100-51, 21-2120-0100-95, 21-2120-0102-02, 21-2120-0102-03, 21-2120-0102-06, 21-2120-0102-07, 21-2120-0102-08, 21-2120-0102-12, 21-2120-0102-13, 21-2120-0102-14, 21-2120-0102-15, 21-2120-0102-17, 21-2120-0102-247, 21-2120-0102-249, 21-2120-0102-51, 21-2120-0102-78, 21-2120-0102-92, 21-2120-0102-98, 21-2120-0103-01, 21-2120-0103-02, 21-2120-0103-03, 21-2120-0103-06, 21-2120-0103-07, 21-2120-0103-08, 21-2120-0103-14, 21-2120-0103-15, 21-2120-0103-17, 21-2120-0103-249, 21-2120-0103-51, 21-2120-0103-78, 21-2120-0104-01, 21-2120-0105-00, 21-2120-0105-01, 21-2120-0105-02L, 21-2120-0105-03L, 21-2120-0105-06L, 21-2120-0105-07L, 21-2120-0105-08L, 21-2120-0105-12L, 21-2120-0105-13L, 21-2120-0105-14L, 21-2120-0105-15L, 21-2120-0105-17L, 21-2120-0105-231, 21-2120-0105-50, 21-2120-0105-50L, 21-2125-0104-01, 21-2125-0105-07L, 21-2125-0105-08L, 21-2125-0105-50L, 21-2127-0104-01, 21-2127-0105-01, 21-2127-0105-02L, 21-2127-0105-03L, 21-2127-0105-06L, 21-2127-0105-08L, 21-2127-0105-50L, 22-4095-0203-51, 22-4095-0204-25, 22-4095-0205-25.

Z-0222-2025
Recall number
Z-0222-2025
Initiated
July 16, 2024
Classification
Class I
Status
Ongoing
Recalling firm
Smiths Medical ASD Inc.
Quantity
138039 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Damage to the battery pack may have caused a short to a capacitor within the battery pack. While the battery encasement is designed to be flame retardant, a short to the capacitor could lead to melting of the battery pack case. If this issue occurs, the battery pack charging circuit may become inoperable.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Component design/selection

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Damage to the battery pack may have caused a short to a capacitor within the battery pack. While the battery encasement is designed to be flame retardant, a short to the capacitor could lead to melting of the battery pack case. If this issue occurs, the battery pack charging circuit may become inoperable.

Code information

All serial numbers.

Distribution pattern

Worldwide Distribution