device · product 1 of 1
LINK Universal Handle, with quick coupling, Stainless Steel, straight, Item Number 130-394/01
Some lots of the LINK Universal Handle, with quick coupling, Stainless Steel, straight may suffer an immediate material failure intraoperatively when the instrument is used for the first time. The instrument can break in the middle.
These labels are deterministic app interpretations, not FDA categories.
Official device-enrichment evidence · Unknown
An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.
Some lots of the LINK Universal Handle, with quick coupling, Stainless Steel, straight may suffer an immediate material failure intraoperatively when the instrument is used for the first time. The instrument can break in the middle.
Code information
UDI-DI: 04026575215539; Lot Codes: C204156; C219063; C219064; C233033; C233034; C332135
Distribution pattern
US Nationwide distribution in the states of FL, KS, MI, NJ, VA.