Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 95398

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
September 12, 2024
Product types
Device
Classifications
Class I
Statuses
Ongoing
Recalling firm wording
Baxter Healthcare Corporation

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Welch Allyn Life2000 Ventilation System, REF MS-01-0118, within exterior carton codes: 1) BT-20-0002, 2) BT-20-0002A, 3) BT-20-0002AP, 4) BT-20-0007, 5) BT-20-0011, and 6) RMS010118CP

Z-0046-2025
Recall number
Z-0046-2025
Initiated
September 12, 2024
Classification
Class I
Status
Ongoing
Quantity
10,957 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The Life2000 ventilator may fail to initiate the Low Gas Pressure alarm if the pressure gas source (Life2000 compressor, oxygen cylinder or wall source) is not supplied to the ventilator before initiating therapy.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software Design Change

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The Life2000 ventilator may fail to initiate the Low Gas Pressure alarm if the pressure gas source (Life2000 compressor, oxygen cylinder or wall source) is not supplied to the ventilator before initiating therapy.

Code information

UDI/DI 00887761978089 or 00815410020537, All Life2000 Ventilators with software version 06.08.00.00

Distribution pattern

US nationwide distribution.