Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 95429

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
September 13, 2024
Product types
Device
Classifications
Class III
Statuses
Ongoing
Recalling firm wording
DRG International, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

17-OH Progesterone ELISA 96 well kit- An enzyme immunoassay for the quantitative in vitro diagnostic measurement of 17-OH Progesterone (17-OHP) in serum or plasma Model/Catalog Number: EIA-1292

Z-0258-2025
Recall number
Z-0258-2025
Initiated
September 13, 2024
Classification
Class III
Status
Ongoing
Recalling firm
DRG International, Inc.
Quantity
504 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The High and Low Control acceptance ranges in the Certificate of Analysis (CofA) did not match the acceptance range information on the High and Low Control Component labels. Due to the defect, the device failed to provide results, which may result in a delay in obtaining test results and a remote chance of an adverse health risk.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process change control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The High and Low Control acceptance ranges in the Certificate of Analysis (CofA) did not match the acceptance range information on the High and Low Control Component labels. Due to the defect, the device failed to provide results, which may result in a delay in obtaining test results and a remote chance of an adverse health risk.

Code information

Lot Code: 04048474012927 (DRG Instruments GmbH) lots: 22K024 and 22K034

Distribution pattern

AZ, CA, NH Foreign: Austria Belgium Georgia Germany Greece India Irak Italy Netherlands Poland Romania Russia Singapore Spain Tunesia Vietnam