Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 95472

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
September 17, 2024
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Zimmer, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Saphena Medical Venapax XL- Endoscopic Vessel Harvesting System for use in minimally invasive surgery allowing access for vessel harvesting and is primarily indicated for patients undergoing endoscopic surgery for arterial bypass. Item Number: VPX4000 (7273)

Z-0285-2025
Recall number
Z-0285-2025
Initiated
September 17, 2024
Classification
Class II
Status
Ongoing
Recalling firm
Zimmer, Inc.
Quantity
432 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
A Gen 5 tail piece was assembled to a Gen 4 Finished Device, resulting in a 0.178" difference in the end of the endoscope slot to the end of the tail piece. This allows the scope to move further into the plastic sheath that runs the length of the device, resulting in an interference fit at the cone, may caus an extension in surgery

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Component change control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

A Gen 5 tail piece was assembled to a Gen 4 Finished Device, resulting in a 0.178" difference in the end of the endoscope slot to the end of the tail piece. This allows the scope to move further into the plastic sheath that runs the length of the device, resulting in an interference fit at the cone, may caus an extension in surgery

Code information

UDI: +M718VPX40001/$$326051543DD6738Q, +M718VPX4001/$$326062143DD8092J Lot Numbers: 43DD6738, 43DD8092

Distribution pattern

Nationwide