Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 95632

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
October 09, 2024
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Shimadzu Corp. Analytical & Measuring Instruments Div.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Autosampler, Models: SIL-20ACXR CL, SIL-20ACHT CL, SIL-30AC CL

Z-0553-2025
Recall number
Z-0553-2025
Initiated
October 09, 2024
Classification
Class II
Status
Ongoing
Quantity
179

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
There have been cases of failure of the main board due to condensation water drained from the Peltier dehumidifier, leaked out from the tray, resulting in a short circuit. In the serious case, the main board and adjacent cables are partly charred. In the minor case, the Y motor and its surrounding die cast part become rusty. These failures could result in delay in patient sample testing.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

There have been cases of failure of the main board due to condensation water drained from the Peltier dehumidifier, leaked out from the tray, resulting in a short circuit. In the serious case, the main board and adjacent cables are partly charred. In the minor case, the Y motor and its surrounding die cast part become rusty. These failures could result in delay in patient sample testing.

Code information

Model/UDI-DI/Serial Number: SIL-20ACXR CL/04540217420131/L20995250001, L20995250002, L20995250003, L20995250004, L20995250005, L20995250006, L20995250007, L20995250008, L20995350010, L20995350011, L20995350012, L20995350013, L20995350014, L20995350015, L20995450016, L20995450017, L20995450018, L20995450019, L20995450020, L20995450021, L20995450022, L20995450023, L20995450024, L20995450025, L20995450026, L20995550027, L20995550028, L20995550029, L20995550030, L20995650031, L20995650032, L20995650033, L20995650034, L20995650035, L20995650036, L20995650037, L20995650038, L20995650039, L20995650040, L20995650041, L20995650042, L20995650043, L20995650044, L20995650045, L20995650046, L20995560001, L20995660002, L20995660003, L20995660004, L20995660005, L20995660006, L20995660007, L20995660008, L20995660009, L20995660010, L20995660011, L20995660012, L20995660013, L20995660014, L20995660015, L20995660016, L20995660017, L20995660018, L20995660019, L20995660020, L20995660021, L20995660022, L20995660023, L20995660024, L20995660025, L20995660026, L20995660027, L20995660028, L20995660029, L20995660030, L20995660031, L20995660032, L20995660033, L20995660034, L20995660035, L20995660036, L20995660037, L20995660038, L20995660039, L20995660040, L20995660041, L20995660042, L20995660043, L20995660044, L20995660045, L20995660046, L20995660047, L20995660048, L20995660049, L20995660050, L20995660051, L20995660052, L20995660053, L20995660054, L20995660055, L20995660056, L20995660057, L20995250009, L20995270001, L20995270002, L20995270003, L20995270004, L20995270005, L20995370006, L20995370007, L20995370008, L20995370009, L20995370010, L20995370011, L20995370012, L20995370013, L20995370014, L20995370015, L20995370016, L20995370017, L20995370018, L20995370019, L20995370020, L20995370021, L20995370022, L20995370023, L20995370024, L20995370025, L20995470028, L20995770032. SIL-20ACHT CL/04540217420124/L20985350001, L20985450002, L20985750003, L20985750004. SIL-30AC CL/04540217420155/L21005250001, L21005250002, L21005250003, L21005250004, L21005250005, L21005250006, L21005250007, L21005350008, L21005350009, L21005350010, L21005350011, L21005350012, L21005350013, L21005350014, L21005350015, L21005350016, L21005350017, L21005350018, L21005350019, L21005350020, L21005350021, L21005350022, L21005350023, L21005350024, L21005450025, L21005550026, L21005550027, L21005650028, L21005650029, L21005750030, L21005750031, L21005570001, L21005570002, L21005670003, L21005670004, L21005670005, L21005670006, L21005670007, L21005670008, L21005770009, L21005770010, L21005770011, L21005770012, L21005770013, L21005770014.

Distribution pattern

Worldwide - US Nationwide distribution in the states of MS, AR, TX, CA, MA, MI, FL, OK, WI, IN, MN, OH, NJ, NY, IL, KY, NV, HI, SD, ID, MI, UT, WA, MO, VA, GA and the countries of DE, CN, JP, UY.