device · product 1 of 1
Atellica CI Analyzer. Catalog Numbers: 10947347.
- Recall number
- Z-0597-2025
- Initiated
- November 04, 2024
- Classification
- Class II
- Status
- Ongoing
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Quantity
- 559 units
App-derived interpretation
Potential that the IMT Diluent volume remaining (% remaining) does not decrease as expected on the Atellica CI Analyzer, potentially leading to the IMT Diluent being empty while still displaying that volume is remaining. In this case Sodium (Na), Potassium (K) and Chloride (Cl) test results may be falsely elevated. Quality Control materials demonstrate the same behavior.
Official device-enrichment evidence · Sourced
Software design
Inspect official wording and provenance
Reason for recall
Potential that the IMT Diluent volume remaining (% remaining) does not decrease as expected on the Atellica CI Analyzer, potentially leading to the IMT Diluent being empty while still displaying that volume is remaining. In this case Sodium (Na), Potassium (K) and Chloride (Cl) test results may be falsely elevated. Quality Control materials demonstrate the same behavior.
Code information
UDI-DI Number: 630414229560; Catalog Number: 10947347. Software version 1.28.51
Distribution pattern
Worldwide distribution - US Nationwide and the countries of Austria, Bangladesh, Bulgaria, Canada, Chile, Colombia, Croatia, Czech Republic, Ecuador, Egypt, France, Germany, Greece, Hong Kong, Hungary, India, Iraq, Israel, Italy, Japan, Jordan, Kenya, Kuwait, Latvia, Malaysia, Netherlands, Oman, Pakistan, Paraguay, Peru, Philippines, Portugal, Qatar, Republic Korea, Romania, Saudi Arabia, Serbia, Slovakia, South Africa, Spain, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, United Kingdom Great Britain, Vietnam.