Exact FDA event identifier

Event 95644

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
November 04, 2024
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Siemens Healthcare Diagnostics, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Atellica CI Analyzer. Catalog Numbers: 10947347.

Z-0597-2025
Recall number
Z-0597-2025
Initiated
November 04, 2024
Classification
Class II
Status
Ongoing
Quantity
559 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Potential that the IMT Diluent volume remaining (% remaining) does not decrease as expected on the Atellica CI Analyzer, potentially leading to the IMT Diluent being empty while still displaying that volume is remaining. In this case Sodium (Na), Potassium (K) and Chloride (Cl) test results may be falsely elevated. Quality Control materials demonstrate the same behavior.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Potential that the IMT Diluent volume remaining (% remaining) does not decrease as expected on the Atellica CI Analyzer, potentially leading to the IMT Diluent being empty while still displaying that volume is remaining. In this case Sodium (Na), Potassium (K) and Chloride (Cl) test results may be falsely elevated. Quality Control materials demonstrate the same behavior.

Code information

UDI-DI Number: 630414229560; Catalog Number: 10947347. Software version 1.28.51

Distribution pattern

Worldwide distribution - US Nationwide and the countries of Austria, Bangladesh, Bulgaria, Canada, Chile, Colombia, Croatia, Czech Republic, Ecuador, Egypt, France, Germany, Greece, Hong Kong, Hungary, India, Iraq, Israel, Italy, Japan, Jordan, Kenya, Kuwait, Latvia, Malaysia, Netherlands, Oman, Pakistan, Paraguay, Peru, Philippines, Portugal, Qatar, Republic Korea, Romania, Saudi Arabia, Serbia, Slovakia, South Africa, Spain, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, United Kingdom Great Britain, Vietnam.