openFDA Device Enforcement
Report-date coverage
June 20, 2012–July 08, 2026
Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
openFDA Device Recall
event_date_initiated coverage
June 01, 1997–June 30, 2026
Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.
Product dates, classifications, firms, and source wording remain attached to their individual rows.
Opening this dossier never hides products that did not match a prior timeline filter.
device · product 1 of 7
Kirwan Surgical Products LLC Disposable 25-gauge bipolar pencil, straight. Model/Catalog Number: 14-5011 The Disposable Microsurgical Bipolar Pencil are designed for use in eye surgery, and other soft tissue procedures. For instance, the pencils are designed for cauterization of the retinal area during surgical procedures. The current line of pencils all have the same molded handle. The pencils' tip outer diameter is 25 gauge.
Potential for the outer stainless-steel tube to break or detach.
Code information
Model/Catalog Number: 14-5011. GTINs: Pouch - 00841019101049, Inner Pack - 10841019101046, Case - 20841019101043. Sterile Lot Range: 220906 to 241002. All expiration dates from 2027-09-01(YYYY-MM_DD) on would encompass the for the recall.
Distribution pattern
Worldwide - US Nationwide including PUERTO RICO; AUSTRALIA, CANADA, CHILE, COLOMBIA, DENMARK, EGYPT, ENGLAND, FRANCE, GERMANY, ISRAEL, ITALY, JAPAN, MEXICO, NETHERLANDS, PHILIPPINES, SOUTH AFRICA, SWITZERLAND, TAIWAN, THAILAND, TRINIDAD, TURKEY, UNITED ARAB EMIRATES, and VENEZUELA.
device · product 2 of 7
Kirwan Surgical Products LLC Disposable 25-gauge bipolar pencil, straight, Long. Model/Catalog Number: 14-5011L The Disposable Microsurgical Bipolar Pencil are designed for use in eye surgery, and other soft tissue procedures. For instance, the pencils are designed for cauterization of the retinal area during surgical procedures. The current line of pencils all have the same molded handle. The pencils' tip outer diameter is 25 gauge.
Potential for the outer stainless-steel tube to break or detach.
Code information
Model/Catalog Number: 14-5011L. GTINs: Pouch - 00841019101056, Inner Pack - 10841019101053, Case - 20841019101050. Sterile Lot Number: 221020. All expiration dates from 2027-09-01(YYYY-MM_DD) on would encompass the for the recall.
Distribution pattern
Worldwide - US Nationwide including PUERTO RICO; AUSTRALIA, CANADA, CHILE, COLOMBIA, DENMARK, EGYPT, ENGLAND, FRANCE, GERMANY, ISRAEL, ITALY, JAPAN, MEXICO, NETHERLANDS, PHILIPPINES, SOUTH AFRICA, SWITZERLAND, TAIWAN, THAILAND, TRINIDAD, TURKEY, UNITED ARAB EMIRATES, and VENEZUELA.
device · product 3 of 7
Accutome/Microsurgical Disposable Bipolar Pencil 5 in. (12.7cm) 25 Gauge Straight. Model/Catalog Number: AB-7004 The Disposable Microsurgical Bipolar Pencil are designed for use in eye surgery, and other soft tissue procedures. For instance, the pencils are designed for cauterization of the retinal area during surgical procedures. The current line of pencils all have the same molded handle. The pencils' tip outer diameter is 25 gauge.
Potential for the outer stainless-steel tube to break or detach.
Code information
Model/Catalog Number: AB-7004. GTINs: Pouch - 00841019113875, Inner Pack - 10841019113872, Case - 20841019113879. All expiration dates from 2027-09-01(YYYY-MM_DD) on would encompass the for the recall.
Distribution pattern
Worldwide - US Nationwide including PUERTO RICO; AUSTRALIA, CANADA, CHILE, COLOMBIA, DENMARK, EGYPT, ENGLAND, FRANCE, GERMANY, ISRAEL, ITALY, JAPAN, MEXICO, NETHERLANDS, PHILIPPINES, SOUTH AFRICA, SWITZERLAND, TAIWAN, THAILAND, TRINIDAD, TURKEY, UNITED ARAB EMIRATES, and VENEZUELA.
device · product 4 of 7
Ambler Surgical Bipolar Pencil 25 Gauge Straight. Model/Catalog Number: PD7226. The Disposable Microsurgical Bipolar Pencil are designed for use in eye surgery, and other soft tissue procedures. For instance, the pencils are designed for cauterization of the retinal area during surgical procedures. The current line of pencils all have the same molded handle. The pencils' tip outer diameter is 25 gauge. Component: No
Potential for the outer stainless-steel tube to break or detach.
Code information
Model/Catalog Number: PD7226. GTINs: Pouch - 00190660001025, Inner Pack - 10190660001022. All expiration dates from 2027-09-01(YYYY-MM_DD) on would encompass the for the recall.
Distribution pattern
Worldwide - US Nationwide including PUERTO RICO; AUSTRALIA, CANADA, CHILE, COLOMBIA, DENMARK, EGYPT, ENGLAND, FRANCE, GERMANY, ISRAEL, ITALY, JAPAN, MEXICO, NETHERLANDS, PHILIPPINES, SOUTH AFRICA, SWITZERLAND, TAIWAN, THAILAND, TRINIDAD, TURKEY, UNITED ARAB EMIRATES, and VENEZUELA.
device · product 5 of 7
Walcott Bipolar Pencil 25 Gauge Straight. Model/Catalog Number: RX14-5011. Product Description: The Disposable Microsurgical Bipolar Pencil are designed for use in eye surgery, and other soft tissue procedures. For instance, the pencils are designed for cauterization of the retinal area during surgical procedures. The current line of pencils all have the same molded handle. The pencils' tip outer diameter is 25 gauge.
Potential for the outer stainless-steel tube to break or detach.
Code information
Model/Catalog Number: RX14-5011. GTINs: Pouch - 00841019109922, Inner Pack - 10841019109929, Case - 20841019109926. All expiration dates from 2027-09-01(YYYY-MM_DD) on would encompass the for the recall.
Distribution pattern
Worldwide - US Nationwide including PUERTO RICO; AUSTRALIA, CANADA, CHILE, COLOMBIA, DENMARK, EGYPT, ENGLAND, FRANCE, GERMANY, ISRAEL, ITALY, JAPAN, MEXICO, NETHERLANDS, PHILIPPINES, SOUTH AFRICA, SWITZERLAND, TAIWAN, THAILAND, TRINIDAD, TURKEY, UNITED ARAB EMIRATES, and VENEZUELA.
device · product 6 of 7
Nidek Co. Disposable Bipolar Pencil 25 Gauge Straight. Model/Catalog Number: 18241-0979. Product Description: The Disposable Microsurgical Bipolar Pencil are designed for use in eye surgery, and other soft tissue procedures. For instance, the pencils are designed for cauterization of the retinal area during surgical procedures. The current line of pencils all have the same molded handle. The pencils' tip outer diameter is 25 gauge.
Potential for the outer stainless-steel tube to break or detach.
Code information
Model/Catalog Number: 18241-0979. GTINs: Pouch - 00841019116623, Inner Pack - 10841019116620, Case - 20841019116627. All expiration dates from 2027-09-01(YYYY-MM_DD) on would encompass the for the recall.
Distribution pattern
Worldwide - US Nationwide including PUERTO RICO; AUSTRALIA, CANADA, CHILE, COLOMBIA, DENMARK, EGYPT, ENGLAND, FRANCE, GERMANY, ISRAEL, ITALY, JAPAN, MEXICO, NETHERLANDS, PHILIPPINES, SOUTH AFRICA, SWITZERLAND, TAIWAN, THAILAND, TRINIDAD, TURKEY, UNITED ARAB EMIRATES, and VENEZUELA.
device · product 7 of 7
Optikon Bipolar Pencil 25 Gauge Straight. Model/Catalog Number: OPT14-5011. Product Description: The Disposable Microsurgical Bipolar Pencil are designed for use in eye surgery, and other soft tissue procedures. For instance, the pencils are designed for cauterization of the retinal area during surgical procedures. The current line of pencils all have the same molded handle. The pencils' tip outer diameter is 25 gauge.
Potential for the outer stainless-steel tube to break or detach.
Code information
Model/Catalog Number: OPT14-5011. Lot Code: J20499-01, J21637-01, J21999-01, J21995-01, J22003-01, J23738-01, J23739-01, J23740-01, J23743-01, J24697-01, J24710-01, J24711-01, J26812-01, J30100-01. Component, no UDI held by Kirwan Surgical Products.
Distribution pattern
Worldwide - US Nationwide including PUERTO RICO; AUSTRALIA, CANADA, CHILE, COLOMBIA, DENMARK, EGYPT, ENGLAND, FRANCE, GERMANY, ISRAEL, ITALY, JAPAN, MEXICO, NETHERLANDS, PHILIPPINES, SOUTH AFRICA, SWITZERLAND, TAIWAN, THAILAND, TRINIDAD, TURKEY, UNITED ARAB EMIRATES, and VENEZUELA.