Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 95914

2 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
February 15, 2024
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Augustine Temperature Management, LLC

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

2 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 2

Augustine Surgical Inc. HotDog Return Electrode accessory connector cables, Model A136 used in conjunction with Warming Blankets and Mattresses.

Z-1171-2025
Recall number
Z-1171-2025
Initiated
February 15, 2024
Classification
Class II
Status
Ongoing
Quantity
20 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Users were experiencing the following error message (M-0B-125) stated Check Neutral Electrode Contact .

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Users were experiencing the following error message (M-0B-125) stated Check Neutral Electrode Contact .

Code information

UDI-DI 00855913001794, Lot Number 2122

Distribution pattern

US distribution to states of: AL, TX, NV

device · product 2 of 2

Augustine Surgical Inc. HotDog Return Electrode accessory connector cables, Model A138 used in conjunction with Warming Blankets and Mattresses.

Z-1172-2025
Recall number
Z-1172-2025
Initiated
February 15, 2024
Classification
Class II
Status
Ongoing
Quantity
7 unit

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Users were experiencing the following error message (M-0B-125) stated Check Neutral Electrode Contact .

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Users were experiencing the following error message (M-0B-125) stated Check Neutral Electrode Contact .

Code information

UDI-DI 00850011479674, Lot Number 2339

Distribution pattern

US distribution to states of: AL, TX, NV