Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 95988

2 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
December 16, 2024
Product types
Device
Classifications
Class I
Statuses
Ongoing
Recalling firm wording
MEDLINE INDUSTRIES, LP - Northfield

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

2 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 2

Medline Fluid Delivery Set with Drip Chamber: 1) FDS,135,CHMN,W/F,MLL,RC(2),PG, Item Number 99000069; 2) FDS,135,NV,MICRO-CHAMBER,W/FILTER,BLUE, Item Number 949000222; 3) FDS,135,NV,MICRO-CHAMBER,W/FILTER,GREEN, Item Number 949000224; 4) FDS,135,NV,MICRO-CHAMBER,W/FILTER,RED, Item Number 949000221; 5) FDS,135,NV,MICRO-CHAMBER,W/FILTER,YELLOW, Item Number 949000223

Z-0764-2025
Recall number
Z-0764-2025
Initiated
December 16, 2024
Classification
Class I
Status
Ongoing
Quantity
13,600 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Fluid Delivery Sets were incorrectly assembled with a white macro drip chamber instead of the required grey micro drip chamber. The macro drip chamber delivers three times more fluid per drop than the micro drip chamber.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Fluid Delivery Sets were incorrectly assembled with a white macro drip chamber instead of the required grey micro drip chamber. The macro drip chamber delivers three times more fluid per drop than the micro drip chamber.

Code information

1)99000069UDI-D:10193489069587(each)30193489069581(case), Lot Number:0000128741; 2)99000069UDI-D:10193489069587(each)30193489069581(case), Lot Number:0000134505; 3)99000069UDI-D:10193489069587(each)30193489069581(case), Lot Number:0000133893; 4)99000069UDI-D:10193489069587(each)30193489069581(case), Lot Number:0000135769; 5)99000069UDI-D:10193489069587(each)30193489069581(case), Lot Number:0000139596; 6)949000222UDI-D:10197344018550(each)30197344018554(case), Lot Number:0000133018; 7)949000224UDI-D:10197344018642(each)30197344018646(case), Lot Number:0000136573; 8)949000221UDI-D:10197344018567(each)30197344018561(case), Lot Number:0000133036; 9)949000223UDI-D:10197344018543(each)30197344018547(case), Lot Number:0000133031;

Distribution pattern

US Nationwide Distribution

device · product 2 of 2

Medline Fluid Delivery Set with Drip Chamber: 1) MTO,135,YELLOW,MICRO,SPK-MAIMONIDES, Item Number: 60130527; 2) MTO,ANGIO,DOUBLE,SPIKE,KIT,KENNEDY HLTH, Item Number: 65111341; 3) MTO,ANGIO,SINGLE,SPIKE,KIT, Item Number: 65232486; 4) MTO,ANGIO,SINGLE,SPIKE,KIT,GOOD,SAM, Item Number: 65071193; 5) MTO,NEURO,FLUSH,SPIKE,KIT,BLUE,FLORIDA, Item Number: 600604422; 6) MTO,NEURO,KIT,-,JERSEY,SHORE PG, Item Number: 60100045; 7) MTO,RADIOLOGY,KIT,HAMOT MED CEN, Item Number: 650801715; 8) MTS,135,CDS,IV,TUBE-JEFFERSON UNIV., Item Number: 60232481; 9) MTS,135,YELLOW,MICRO,SPIKE,LUTHERAN MED, Item Number: 60120583; 10) MTS,ANGIO,NEURO,KIT,NORTH SHORE UNIV, Item Number: 601403920; 11) MTS,CUST. FILTER-NEWYORK UNIVERSITY, Item Number: 62140602; 12) MTS,INTERV,NEURO,KIT-WESTCHESTER,MC, Item Number: 602301614; 13) MTS,NEU,NORTH SHORE HOSPITAL, Item Number: 6010771601; 14) MTS,NEURO,KIT, Item Number: 600602914; 15) MTS,NEURO,KIT,2,NEWENGLAND MED CTR, Item Number: 601400111; 16) MTS,NEURO,KIT,YLW&RED,FLORIDA HOSP, Item Number: 600604423; 17) MTS,NEURO,KIT-JOHN,F,KENNEDY MED CENTER, Item Number: 60110552; 18) MTS,NEURO,KIT-OVERLOOK HOSPITAL PG, Item Number: 60150035; 19) MTS,NEURO,SPIKE,KIT,CAPITAL HLTHSYS-FULD, Item Number: 60182723; 20) MTS,NEURO-RADIOLOGY,KIT-YALE-NEWHAVEN, Item Number: 602500214; 21) MTS,STROKE/EMBO,KIT,NORTH SHORE UNIV, Item Number: 601403921; 22) MTS,YELLOW,/,RED,MULTILINE,KIT,MAIMONIDE, Item Number: 601305210; 23) MTS,YELLOW,/,RED,MULTILINE,KIT,MAIMONIDE, Item Number: 601400110;

Z-0765-2025
Recall number
Z-0765-2025
Initiated
December 16, 2024
Classification
Class I
Status
Ongoing
Quantity
3930 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Fluid Delivery Sets were incorrectly assembled with a white macro drip chamber instead of the required grey micro drip chamber. The macro drip chamber delivers three times more fluid per drop than the micro drip chamber.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Fluid Delivery Sets were incorrectly assembled with a white macro drip chamber instead of the required grey micro drip chamber. The macro drip chamber delivers three times more fluid per drop than the micro drip chamber.

Code information

1) 60130527 UDI-D:10193489046465(each), 20193489046462(case), Lot Number:0000133032; 2) 65111341 UDI-D:10193489061314(each), 20193489061311(case), Lot Number:0000129541; 3) 65111341 UDI-D:10193489061314(each), 20193489061311(case), Lot Number:0000134701; 4) 65232486 UDI-D:10193489063950(each), 20193489063957(case), Lot Number:0000129544; 5) 65232486 UDI-D:10193489063950(each), 20193489063957(case), Lot Number:0000128715; 6) 65232486 UDI-D:10193489063950(each), 20193489063957(case), Lot Number:0000133060; 7) 65232486 UDI-D:10193489063950(each), 20193489063957(case), Lot Number:0000131657; 8) 65232486 UDI-D:10193489063950(each), 20193489063957(case), Lot Number:0000135723; 9) 65232486 UDI-D:10193489063950(each), 20193489063957(case), Lot Number:0000135294; 10) 65071193 UDI-D:10193489060904(each), 20193489060901(case), Lot Number:0000130171; 11) 600604422 UDI-D:10193489070224(each), 20193489070221(case), Lot Number:0000135606; 12) 600604422 UDI-D:10193489070224(each), 20193489070221(case), Lot Number:0000134729; 13) 600604422 UDI-D:10193489070224(each), 20193489070221(case), Lot Number:0000137115; 14) 600604422 UDI-D:10193489070224(each), 20193489070221(case), Lot Number:0000130833; 15) 600604422 UDI-D:10193489070224(each), 20193489070221(case), Lot Number:0000131669; 16) 600604422 UDI-D:10193489070224(each), 20193489070221(case), Lot Number:0000129513; 17) 600604422 UDI-D:10193489070224(each), 20193489070221(case), Lot Number:0000128721; 18) 60100045 UDI-D:10193489044928(each), 30193489044922(case), Lot Number:0000129523; 19) 60100045 UDI-D:10193489044928(each), 30193489044922(case), Lot Number:0000132393; 20) 60100045 UDI-D:10193489044928(each), 30193489044922(case), Lot Number:0000134940; 21) 60100045 UDI-D:10193489044928(each), 30193489044922(case), Lot Number:0000144158; 22) 60100045 UDI-D:10193489044928(each), 30193489044922(case), Lot Number:0000139082; 23) 650801715 UDI-D:10193489075458(each), 20193489075455(case), Lot Number:0000130138; 24) 650801715 UDI-D:10193489075458(each), 20193489075455(case), Lot Number:0000132350; 25) 650801715 UDI-D:10193489075458(each), 20193489075455(case), Lot Number:0000135736; 26) 60232481 UDI-D:10193489053326(each), 30193489053320(case), Lot Number:0000129534; 27) 60120583 UDI-D:10193489045864(each), 30193489045868(case), Lot Number:0000133660; 28) 60120583 UDI-D:10193489045864(each), 30193489045868(case), Lot Number:0000136501; 29) 60120583 UDI-D:10193489045864(each), 30193489045868(case), Lot Number:0000131600; 30) 601403920 UDI-D:10193489071498(each), 30193489071492(case), Lot Number:0000131376; 31) 601403920 UDI-D:10193489071498(each), 30193489071492(case), Lot Number:0000130940; 32) 62140602 UDI-D:10193489057805(each), 30193489057809(case), Lot Number:0000132222; 33) 62140602 UDI-D:10193489057805(each), 30193489057809(case), Lot Number:0000141215; 34) 602301614 UDI-D:10193489072754(each), 30193489072758(case), Lot Number:0000133659; 35) 6010771601 UDI-D:10193489099843(each), 30193489099847(case), Lot Number:0000127859; 36) 600602914 UDI-D:10193489070170(each), 30193489070174(case), Lot Number:0000135717; 37) 600602914 UDI-D:10193489070170(each), 30193489070174(case), Lot Number:0000128447; 38) 601400111 UDI-D:10193489071344(each), 30193489071348(case), Lot Number:0000133121; 39) 601400111 UDI-D:10193489071344(each), 30193489071348(case), Lot Number:0000130573; 40) 600604423 UDI-D:10193489070231(each), 30193489070235(case), Lot Number:0000128607; 41) 600604423 UDI-D:10193489070231(each), 30193489070235(case), Lot Number:0000137751; 42) 60110552 UDI-D:10193489045611(each), 30193489045615(case), Lot Number:0000131878; 43) 60110552 UDI-D:10193489045611(each), 30193489045615(case), Lot Number:0000130577; 44) 60110552 UDI-D:10193489045611(each), 30193489045615(case), Lot Number:0000138084; 45) 60150035 UDI-D:10193489048292(each), 30193489048296(case), Lot Number:0000130187; 46) 60150035 UDI-D:10193489048292(each), 30193489048296(case), Lot Number:0000131883; 47) 60150035 UDI-D:10193489048292(each), 30193489048296(case), Lot Number:0000143095; 48) 60182723 UDI-D:10193489049817(each), 30193489049811(case), Lot Number:0000130988; 49) 602500214 UDI-D:10193489072815(each), 30193489072819(case), Lot Number:0000127874; 50) 601403921 UDI-D:30193489071508(each), 30193489071508(case), Lot Number:0000131619; 51) 601403921 UDI-D:30193489071508(each), 30193489071508(case), Lot Number:0000133025; 52) 601403921 UDI-D:30193489071508(each), 30193489071508(case), Lot Number:0000136559; 53) 601305210 UDI-D:10193489070989(each), 30193489070983(case), Lot Number:0000130984; 54) 601400110 UDI-D:10193489071337(each), 30193489071331(case), Lot Number:0000128757;