Baxter NovumIQ Syringe INFUSION SYSTEM , REF 40800BAXUS
- Recall number
- Z-1423-2025
- Initiated
- March 03, 2025
- Classification
- Class II
- Status
- Ongoing
- Recalling firm
- Baxter Healthcare Corporation
- Quantity
- 149 units
App-derived interpretation
Baxter Healthcare Corporation is issuing an Urgent Medical Device Recall for the Novum IQ syringe pumps listed below due to an incorrectly installed gasket. A gasket failure could lead to potential fluid ingress due to cleaning, or as a result of an IV fluid spill.
Official device-enrichment evidence · Sourced
Process control
Inspect official wording and provenance
Reason for recall
Baxter Healthcare Corporation is issuing an Urgent Medical Device Recall for the Novum IQ syringe pumps listed below due to an incorrectly installed gasket. A gasket failure could lead to potential fluid ingress due to cleaning, or as a result of an IV fluid spill.
Code information
UDI/DI 05413765852428, Serial Numbers: 02A220600103, 02A221100387, 02A240800201, 02A230300471, 02A240800161, 02A220600112, 02A230100474, 02A240800343, 02A230400357, 02A240800175, 02A220700259, 02A230200471, 02A240800710, 02A230600165, 02A240800223, 02A220700176, 02A230200487, 02A240900081, 02A230600175, 02A240800513, 02A220700349, 02A230300030, 02A240900133, 02A230600176, 02A240800577, 02A220700268, 02A230300117, 02A240900265, 02A230600255, 02A240900261, 02A220700081, 02A230300125, 02A240900277, 02A230600256, 02A240900421, 02A220700140, 02A230300132, 02A240900285, 02A230600618, 02A240900742, 02A220700351, 02A230300146, 02A240900287, 02A230700137, 02A240900746, 02A220700409, 02A230300275, 02A240900282, 02A241000576, 02A240901067, 02A220700368, 02A230300287, 02A240900373, 02A220700166, 02A240901181, 02A220700428, 02A230300309, 02A240900407, 02A221000108, 02A240901422, 02A220700450, 02A230300350, 02A240900422, 02A221000346, 02A240901428, 02A220700454, 02A230300389, 02A240900423, 02A221000349, 02A240901430, 02A220700458, 02A230300477, 02A240900443, 02A221100154, 02A241000027, 02A220800010, 02A230300520, 02A240900427, 02A221100208, 02A241000051, 02A220800060, 02A230300547, 02A240900502, 02A221100201, 02A241000251, 02A220800201, 02A230400035, 02A240900681, 02A230200462, 02A241000254, 02A220800407, 02A230400145, 02A240900677, 02A230200495, 02A241000366, 02A221000109, 02A230600421, 02A240900749, 02A230300091, 02A241000479, 02A221000233, 02A230600549, 02A240900811, 02A230300342, 02A241000862, 02A221000240, 02A230600792, 02A240900981, 02A230300367, 02A241001284, 02A221000360, 02A230600820, 02A240901095, 02A230300417, 02A241001392, 02A221000099, 02A230600894, 02A240901196, 02A230300533, 02A241001413, 02A221100056, 02A230600908, 02A240901243, 02A230400030, 02A241001478, 02A221100116, 02A230700053, 02A241000498, 02A230600496, 02A241001496, 02A221100244, 02A230700043, 02A241000939, 02A240800026, 02A241100021, 02A221100260, 02A240600468, 02A241000957, 02A240800050, 02A241100037, 02A221100294, 02A240800006, 02A230300466, 02A240800069, 02A241100047, 02A221100346, 02A240800007, 02A230300467, 02A240800142.
Distribution pattern
US Nationwide distribution in the states of AR, CA, FL, IA, IN, LA, MO, MT, NC, NY, OH, OR, PA, TN, TX, VA, WA, WI, WV.