Exact FDA event identifier

Event 96439

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
March 20, 2025
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Maquet Cardiovascular, LLC

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

The VASOVIEW HEMOPRO 2 Endoscopic Vessel Harvesting System is designed for use in conjunction with the 7 mm extended length endoscope. The Harvesting Cannula has four lumens which house the endoscope, C-Ring, distal lens washer tube, and Harvesting Tool for cutting and cauterizing vessel branches. The Harvesting Tool has two curved jaws. One of the jaws contains the heating elements for branch cutting and cauterization and spot cautery. An area near the tip of the convex side of the jaw is used for spot cautery. The activation toggle is used to control the jaws to activate the heating elements. Positioning the device, cutting, and cauterization are performed under endoscopic visualization. The device is intended for specific use with the VASOVIEW HEMOPRO Power Supply, VASOVIEW HEMOPRO 2 extension cable, and an adapter cable.

Z-1622-2025
Recall number
Z-1622-2025
Initiated
March 20, 2025
Classification
Class II
Status
Ongoing
Quantity
22,244 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
There have been at least 12 complaints from users as of March 3, 2025, reporting that they identified the ceramic C-Ring to be broken in half during use of the device. The ceramic C-Ring has a potential to break while in use. Maquet/Getinge determined that the issue was limited to the units of VH-4000 and VH-4001 manufactured with ceramic C-Ring, and the ceramic C-Ring was implemented in the Hemopro 2 devices from November 5, 2024, through February 14, 2025.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

There have been at least 12 complaints from users as of March 3, 2025, reporting that they identified the ceramic C-Ring to be broken in half during use of the device. The ceramic C-Ring has a potential to break while in use. Maquet/Getinge determined that the issue was limited to the units of VH-4000 and VH-4001 manufactured with ceramic C-Ring, and the ceramic C-Ring was implemented in the Hemopro 2 devices from November 5, 2024, through February 14, 2025.

Code information

Model No. VH-4000 (Hemopro 2), VH-4001 (Hemopro 2 w/ Vasoshield); UDI-DI: 00607567700406 (Hemopro 2), 00607567700451 (Hemopro 2 w/ Vasoshield); Lot/Serial No. 3000436447, 3000437191, 3000439557, 3000439600, 3000439623, 3000439922, 3000440187, 3000440220, 3000440283, 3000440754, 3000440957, 3000440981, 3000440982, 3000441144, 3000441319, 3000441541, 3000441774, 3000442133, 3000442263, 3000442323, 3000442627, 3000442871, 3000442953, 3000442955, 3000443198, 3000443216, 3000443545, 3000443632, 3000443851, 3000444131, 3000444151, 3000444484, 3000445702, 3000445704, 3000445941, 3000446003, 3000446365, 3000446489, 3000446709, 3000446774, 3000447195, 3000447535, 3000447597, 3000447845, 3000448208, 3000448471, 3000448761, 3000449047, 3000449170, 3000449267, 3000450191, 3000450967.

Distribution pattern

Worldwide distribution - US Nationwide and the countries of Mexico, United Arab Emirates, Singapore, India, Australia, Canada, Brazil, Hong Kong, Japan, South Korea, New Zealand.