Exact FDA event identifier

Event 96512

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
March 20, 2025
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Beckman Coulter Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

DxC 500 AU Clinical Chemistry Analyzer, Catalog Numbers/UDI codes: C63519 / 14987666545058 C63520 / 14987666545065. The Beckman Coulter DxC 500 AU Clinical Chemistry Analyzer is an automated chemistry analyzer that measures analytes in samples, in combination with appropriate reagents, calibrators, quality control (QC) material and other accessories. This system is for in vitro diagnostic use only.

Z-1784-2025
Recall number
Z-1784-2025
Initiated
March 20, 2025
Classification
Class II
Status
Ongoing
Recalling firm
Beckman Coulter Inc.
Quantity
Unavailable

App-derived interpretation

Device software or design reason.device_software_design · v1.0.0
software error

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Beckman Coulter is recalling their DxC 500 AU Clinical Chemistry Analyzer because a software error causes the analyzer to not run a requested calibration order in the following scenario: when a reagent blank or calibration is ordered during sample processing and then any of the components (R1 and/or R2) depletes to zero tests, the analyzer will not be able to complete the calibration request, and the calibration order will remain pending. No further calibration orders can be processed for any assays, and the instrument refuses to accept further sample processing order after the existing calibration curves are expired. Although in-process tests will be completed, this error can cause a delay in reporting subsequent test results. No further calibration orders can be processed for any assays which may cause a delay in reporting test results.

Code information

Catalog Number: C63519 UDI-DI code: 14987666545058 Serial Numbers: Catalog Number: C63520 UDI-DI code: 14987666545065 Serial Numbers:

Distribution pattern

U.S: AL, CA, GA, ID, IN, LA, MA, MI, MN, MT, NC, NV, PA, PR, TN , TX, and WI O.U.S.: Australia, Belgium, Bosnia and Herzegovina, Canada, Chile, Croatia, Egypt, Ethiopia, France, Germany, Ghana, India, Italy, Korea, Lebanon, Libya, Malaysia, Netherlands, North Macedonia, Oman, Paraguay, Philippines, Poland, Puerto Rico, Qatar, Saudi Arabia, South Africa, Spain, Switzerland, and Tunisia