Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 96539

2 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
March 19, 2025
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Siemens Healthcare Diagnostics Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

2 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 2

epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736515; epoc Software Version & Sensor Configuration: epoc Host SW v3.41.2, epoc NXS SW v4.14.9 & 4.14.11, Sensor Configuration 45.n.

Z-1684-2025
Recall number
Z-1684-2025
Initiated
March 19, 2025
Classification
Class II
Status
Ongoing
Quantity
264,252 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The occurrence of discrepant high pH results in samples introduced with higher injection volumes with epoc sensor configuration 45.n.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The occurrence of discrepant high pH results in samples introduced with higher injection volumes with epoc sensor configuration 45.n.

Code information

Siemens Material Number (SMN): 10736515; Udi-DI: 00809708121860; Lot Numbers: All lot numbers with prefix (xx) 02 or 12; epoc Software Version and Senor Configuration: epoc Host SW V3.41.2, epoc NXS SW v4.14.9 & 4.14.11, Sensor Configuration 45.n.

Distribution pattern

Domestic: Nationwide Distribution; Foreign: Albania, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bolivia, Bosnia, Brazil, Cambodia, Canada, Canary Islands, Chile, China, Colombia, Croatia, Ecuador, Estonia, France, Germany, Ghana, Greece, Guadeloupe, Guatemala, Hong Kong, Hungary, Iceland, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kuwait, Lithuania, Malaysia, Mali, Martinique, Mexico, Morocco, Netherlands, New Caledonia, New Zealand, Nigeria, Oman, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Federation, San Marino, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Vietnam.

device · product 2 of 2

epoc BGEM Crea Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736382; epoc Software Version & Sensor Configuration: epoc Host SW v3.41.2, epoc NXS SW v4.14.9 & 4.14.11, Sensor Configuration 45.n.

Z-1685-2025
Recall number
Z-1685-2025
Initiated
March 19, 2025
Classification
Class II
Status
Ongoing
Quantity
5,207 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The occurrence of discrepant high pH results in samples introduced with higher injection volumes with epoc sensor configuration 45.n.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The occurrence of discrepant high pH results in samples introduced with higher injection volumes with epoc sensor configuration 45.n.

Code information

Siemens Material Number (SMN): 10736382; Udi-DI: 00809708072254; Lot Numbers: All lot numbers with prefix (xx) 00 or 06; epoc Software Version and Senor Configuration: epoc Host SW V3.41.2, epoc NXS SW v4.14.9 & 4.14.11, Sensor Configuration 45.n.

Distribution pattern

Domestic: Nationwide Distribution; Foreign: Albania, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bolivia, Bosnia, Brazil, Cambodia, Canada, Canary Islands, Chile, China, Colombia, Croatia, Ecuador, Estonia, France, Germany, Ghana, Greece, Guadeloupe, Guatemala, Hong Kong, Hungary, Iceland, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kuwait, Lithuania, Malaysia, Mali, Martinique, Mexico, Morocco, Netherlands, New Caledonia, New Zealand, Nigeria, Oman, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Federation, San Marino, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Vietnam.