openFDA Device Enforcement
Report-date coverage
June 20, 2012–July 08, 2026
Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
openFDA Device Recall
event_date_initiated coverage
June 01, 1997–June 30, 2026
Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.
Product dates, classifications, firms, and source wording remain attached to their individual rows.
Opening this dossier never hides products that did not match a prior timeline filter.
The mounting hardware that is supplied with the 4FXS-60 monitor bracket does not provide adequate thread engagement to support the weight of the frame and monitor. This could lead to a failure of the mounting hardware and subsequent separation of the frame from the mounting column.
These labels are deterministic app interpretations, not FDA categories.
The mounting hardware that is supplied with the 4FXS-60 monitor bracket does not provide adequate thread engagement to support the weight of the frame and monitor. This could lead to a failure of the mounting hardware and subsequent separation of the frame from the mounting column.
Code information
UDI/DI 10841736116071, All Serial Numbers
Distribution pattern
US Nationwide distribution including in the states of AL, AZ, CA, FL, IA, ID, IL, LA, MA, MD, MI, MN, NC, NH, NJ, NM, NY, PA, TN, TX, UT, VA, WI, WV.