device · product 1 of 2
Washer Disinfector Aquadis 56; Model Number: 56A
- Recall number
- Z-1720-2025
- Initiated
- April 02, 2025
- Classification
- Class II
- Status
- Ongoing
- Recalling firm
- Getinge Disinfection Ab
- Quantity
- 102 units (8 US, 94 OUS)
App-derived interpretation
The frequency inverter is not programmed by supplier according to specification which causes the circulation pump to start earlier than expected, resulting in additional wear and tear, earlier than expected, on the axial seal in the pump housing.
Official device-enrichment evidence · Unknown
An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.
Inspect official wording and provenance
Reason for recall
The frequency inverter is not programmed by supplier according to specification which causes the circulation pump to start earlier than expected, resulting in additional wear and tear, earlier than expected, on the axial seal in the pump housing.
Code information
Model: 56A; UDI-DI: 07340153710344; Serial Numbers: WAA104857, WAA107934, WAA107937, WAA108359, WAA108540, WAA108963, WAA109338, WAA110115, WAA108201, WAA109175, WAA109884, WAA109989, WAA109998, WAA110804, WAA600035, WAA109641, WAA109651, WAA111118, WAA106820, WAA106835, WAA106841, WAA107693, WAA107428, WAA107429, WAA600055, WAA600060, WAA600061, WAA106607, WAA106668, WAA106753, WAA107358, WAA109804, WAA106793, WAA108419, WAA110679, WAA109022, WAA109030, WAA109564, WAA109566, WAA109595, WAA109598, WAA105283, WAA107346, WAA600188, WAA107314, WAA107330, WAA110062, WAA110075, WAA110084, WAA110104, WAA110163, WAA110302, WAA110201, WAA600558, WAA110953, WAA110963, WAA110138, WAA109975, WAA107281, WAA107303, WAA108972, WAA110228, WAA110162, WAA110727, WAA110738, WAA110742, WAA110803, WAA108669, WAA108670, WAA108673, WAA108722, WAA110762, WAA110631, WAA110636, WAA600167, WAA600168, WAA110482, WAA111159, WAA111170, WAA111183, WAA105284, WAA108615, WAA109514, WAA110174, WAA110178, WAA110898, WAA110907, WAA600077, WAA600177, WAA600338, WAA600339, WAA600340, WAA600341, WAA600342, WAA600343, WAA600344, WAA600394, WAA600705, WAA600706, WAA600711, WAA600712, WAA600713.
Distribution pattern
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Bangladesh, Brazil, Canada, China, Colombia, Costa Rica, Cote d'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Italy, Japan, Kenya, Korea, Kuwait, Latvia, Lithuania, Malaysia, Mauritania, Mexico, Netherlands, New Caledonia, Nigeria, Norway, Oman, Pakistan, Philippines, Poland, Romania, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, Spain, Sweden, Switzerland, Syria, Thailand, Turkey, Turkmenistan, United Arab Emirates, United Kingdom, Vietnam, Zambia.