Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 96692

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
April 09, 2025
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Richard Wolf GmbH

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

PiezoWave 2 Control Unit, Product Number 100506US, compact myofascial acoustic compression therapy device that enables treatment with piezoelectric shockwaves (Therapeutic massager).

Z-1753-2025
Recall number
Z-1753-2025
Initiated
April 09, 2025
Classification
Class II
Status
Ongoing
Recalling firm
Richard Wolf GmbH
Quantity
31 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The high voltage capacitor may fail early in the product life cycle after delivery of surge voltages. When this failure occurs, the capacitor may produce smoke and a burning smell. In extremely rare instances a capacitor may cause a spark or excessive heat when an internal capacitor failure occurs.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Nonconforming Material/Component

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The high voltage capacitor may fail early in the product life cycle after delivery of surge voltages. When this failure occurs, the capacitor may produce smoke and a burning smell. In extremely rare instances a capacitor may cause a spark or excessive heat when an internal capacitor failure occurs.

Code information

UDI-DI: 00814334022863; Serial Numbers: 5000553232 5000553233 5000553234 5000553235 5000553236 5000553237 5000553238 5000553239 5000553240 5000553241 5000553242 5000553243 5000553244 5000553245 5000553246 5000553247 5000553248 5000553249 5000553250 5000553251 5000553482 5000553483 5000553484 5000553485 5000553486 5000553488 5000553502 5000553503 5000553504 5000553505 5000553506

Distribution pattern

US State: GA