device · product 1 of 2
Atellica IM Homocysteine Calibrator (2 Pack). Siemens Material Number: 10995498
- Recall number
- Z-1821-2025
- Initiated
- April 15, 2025
- Classification
- Class II
- Status
- Ongoing
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Quantity
- 2,212 packs (4,424 units) - 448 packs US, 1,764 packs OUS
App-derived interpretation
Certain lots of Atellica IM Homocysteine Calibrator and ADVIA Centaur HCY Calibrator do not meet the assay Instructions for Use (IFU) reference interval due to calibrator instability resulting in increased positive bias.
Official device-enrichment evidence · Unknown
An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.
Inspect official wording and provenance
Reason for recall
Certain lots of Atellica IM Homocysteine Calibrator and ADVIA Centaur HCY Calibrator do not meet the assay Instructions for Use (IFU) reference interval due to calibrator instability resulting in increased positive bias.
Code information
Siemens Material Number: 10995498. UDI-DI: 00630414598253 Kit lot numbers: 86236A74, 44206A75. ***Updated 5/8/26*** Additional Lots 25675A77, 44505A78, 52364A79.
Distribution pattern
Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, IL, IN, KY, MA, MD, ME, MI, MO, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, TX, UT, VA, VT, WI, WY and the countries of Austria, Belgium, Croatia, Curacao,¿St¿Eus, Czech¿Republic, Denmark, Egypt, Finland, France, Georgia, Germany, Greece, Hungary, Iran, Ireland, Israel, Italy, Jordan, Kuwait, Latvia, Lithuania, Nepal, Netherlands, Norway, Pakistan, Poland, Portugal, Romania, Serbia, Slovakia, Spain, Sweden, Switzerland, Turkey, Turkmenistan, U.A.E., United¿Kingdom, Vatikan City, Zaire.