device · product 1 of 1
Olympus EVIS EXERA III Duodenovideoscope. Model Number: TJF-Q190V
- Recall number
- Z-1911-2025
- Initiated
- May 05, 2025
- Classification
- Class II
- Status
- Ongoing
- Recalling firm
- Olympus Corporation of the Americas
- Quantity
- 12,195 units
App-derived interpretation
microbial contamination
Official device-enrichment evidence · Sourced
Labeling design
Inspect official wording and provenance
Reason for recall
Recent post-market surveillance data suggests a possible association of higher microbial contamination levels from TJF duodenoscopes when manual cleaning was delayed beyond one hour and a presoak was performed, compared to those TJF duodenoscopes where manual cleaning began within one hour after patient procedure.
Code information
UDI-DI: 04953170405563. Model Number: TJF-Q190V. All serial numbers.
Distribution pattern
Worldwide Distribution: US (Nationwide) and OUS (International) countries to: Algeria, Angola, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Belarus, Belgium, Benin, Bolivia, Brazil, Canada, Cayman Islands, Chile, Colombia, Costa Rica, Croatia, Denmark, Dominican Republic, Egypt, El Salvador, Estonia, Ethiopia, Faroe Islands, Finland, France, French Guiana, Georgia, Germany, Ghana, Guadeloupe, Guatemala, Honduras, Hong Kong, Iceland, India, Iraq, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Kuwait, Kyrgyzstan, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Malaysia, Malta, Martinique, Mauritius, Mexico, Monaco, Morocco, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Palestinian Authority, Panama, Paraguay, Peru, Poland, Portugal, Qatar, Republic of the Congo, R¿union, Russia, Saudi Arabia, Senegal, Seychelles, Singapore, South Africa, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, Uruguay, Vietnam, Yemen.