Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 96737

5 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
May 12, 2025
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
BD SWITZERLAND SARL

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

5 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 5

Alaris Pump Infusion Set: SmartSite Bag Access Non-Vented Bonded Texium With Priming Cap Back Check Valve 2 SmartSite Y-Sites, REF: 10013361T; SmartSite Bag Access Non-Vented Bonded Texium Closed Male Luer With Priming Cap, REF: 10321213T; SmartSite Bag Access Non-Vented Low Sorbing Tubing (PE Lined) Bonded Texium Closed Male Luer With Priming Cap, REF: 22000-B007T; Low Sorbing Tubing (PE Lined) Bonded TexiumTM with priming cap, REF: 22600-0007T; SmartSite Bag Access Non-Vented Low Sorbing Tubing (PE Lined) Bonded Texium With Priming Cap, REF: 22602-B007T; SmartSite Bag Access Non-Vented Low Sorbing Tubing (PE Lined) Bonded Texium With Priming Cap, REF: 22602-B007T; SmartSite Bag Access Non-Vented Low Sorbing Tubing (PE Lined) Bonded Texium With Priming Cap, REF: 22603-B007T; Bonded Texium Closed Male Luer with priming cap Back Check Valve 2 SmartSite Y-Sites, REF: 24010-0007T; SmartSite Bag Access Non-Vented 0.2 MicronFilter Bonded Texium Closed Low Sorbing Tubing (PE Lined) Male Luer with priming cap SmartSite Y-site, REF: 24301-0007T; SmartSite Bag Access Non-Vented Low Sorbing Texium Back Check Valve 2 SmartSite Y-Sites, REF: 24601-B007T. BD 31-IN SEC SET 20DP W/BAG ACCESS PORT TEXIUMTM AND HANGER LOW SORBING, REF: 70001B-07T; BD 20-IN EXTENSION SET W/0.2 MICRON FILTER ONE SmartSite VALVE AND Texium, REF: 20350ET; BD Secondary Set 20 Drop Vented Non Vented With Texium & Hanger, REF: 40000-07T; BD 20 DROP SECONDARY SET SmartSite VALVE BAG ACCESS W/TEXIUM CLOSED MALE LUER HANGER, REF: 10013364T; BD 32-IN GRAV SET MICRON FILTER BAG ACCESS PORT SS VALVE TEXIUMTM LOW ABSORPTION, REF: 4030B-07T; BD Texium Smartsite MICROBORE EXTENSION SET 60 IN, REF: 22005E-07T

Z-1970-2025
Recall number
Z-1970-2025
Initiated
May 12, 2025
Classification
Class II
Status
Ongoing
Recalling firm
BD SWITZERLAND SARL
Quantity
2,024,780

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Sterile, single use closed system drug transfer devices, standalone or on infusion sets or bonded syringes, may leak at the interface between the closed male luer component and the needle-free connector, which may cause procedure delay, treatment interruptions, exposure to hazardous drugs, and/or underdosing of medication.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Sterile, single use closed system drug transfer devices, standalone or on infusion sets or bonded syringes, may leak at the interface between the closed male luer component and the needle-free connector, which may cause procedure delay, treatment interruptions, exposure to hazardous drugs, and/or underdosing of medication.

Code information

REF/UDI-DI/Lot(Expiration): 10013361T/10885403223204/23025303(February 10, 2026), 23035235(March 11, 2026), 23045223(April 6, 2026), 23075395(July 17, 2026), 23085227(August 8, 2026), 23115150(November 8, 2026), 23115151(November 8, 2026), 23125167(December 5, 2026), 24025684(February 6, 2027), 24045377(April 23, 2027), 24055576(May 27, 2027), 24075224(July 26, 2027), 24105079(October 11, 2027), 24115159(November 7, 2027); 10321213T/10885403228971/23035169(March 15, 2026), 23045174(April 4, 2026), 23085140(August 3, 2026), 23095142(September 11, 2026), 23105297(October 12, 2026), 24025686(February 6, 2027); 22000-B007T/10885403245022/23025320(February 13, 2026), 23045224(April 11, 2026), 23065209(June 10, 2026), 23075394(July 21, 2026), 23095143(September 11, 2026), 23105133(October 6, 2026), 24015278(January 13, 2027), 24025685(February 6, 2027), 24036209(March 25, 2027), 24075260(July 26, 2027), 24066653(June 24, 2027); 22600-0007T/10885403239595/23095151(September 11, 2026), 24036205(March 25, 2027), 24055601(May 27, 2027); 22602-B007T/10885403240935/23045501(April 20, 2026), 23095152(September 11, 2026), 24015186(January 9, 2027), 24026152(February 27, 2027), 24036206(March 25, 2027), 24045384(April 23, 2027), 24115032(November 2, 2027); 22603-B007T/10885403240942/23035482(March 23, 2026), 23035504(March 25, 2026), 23055379(May 23, 2026), 23065210(June 12, 2026), 23075056(July 13, 2026), 23085141(August 3, 2026), 23095808(September 14, 2026), 23105296(October 12, 2026), 23125107(December 4, 2026), 24015185(January 9, 2027), 24025681(February 6, 2027), 24026153(February 27, 2027), 24026154(February 27, 2027), 24036207(March 25, 2027), 24045381(April 23, 2027), 24055577(May 27, 2027), 24066654(June 24, 2027), 24085243(August 26, 2027), 24095223(September 23, 2027), 24115312(December 2, 2027); 24010-0007T/10885403271021/23025108(February 8, 2026), 23035249(March 14, 2026), 23075392(July 18, 2026), 23085223(August 8, 2026), 23095763(September 19, 2026), 23105136(October 6, 2026), 23115054(November 3, 2026), 24015279(January 13, 2027), 24015386(January 17, 2027), 24025683(February 6, 2027), 24026149(February 27, 2027), 24045382(April 23, 2027), 24055598(May 27, 2027), 24075225(July 26, 2027), 24085240(August 26, 2027), 24095220(September 23, 2027), 24115311(November 14, 2027), 24066656(June 24, 2027); 24301-0007T/10885403223198/23025183(February 9, 2026), 23025238(February 10, 2026), 23035338(March 16, 2026), 23045222(April 12, 2026), 23075393(July 20, 2026), 23085226(August 8, 2026), 23095764(September 19, 2026), 23105284(October 11, 2026), 23115055(November 3, 2026), 24026089(February 21, 2027), 24026150(February 27, 2027), 24036178(March 25, 2027), 24045383(April 23, 2027), 24055599(May 27, 2027), 24075212(July 26, 2027), 24085241(August 26, 2027), 24095221(September 23, 2027), 24115279(November 12, 2027), 24015474(January 20, 2027); 24601-B007T/10885403240959/23035170(March 9, 2026), 23055380(May 24, 2026), 23065280(June 13, 2026), 23085142(August 3, 2026), 23095153(September 11, 2026), 23105285(October 11, 2026), 23115052(November 3, 2026), 24025682(February 6, 2027), 24026151(February 27, 2027), 24036208(March 25, 2027), 24055600(May 27, 2027), 24075261(July 26, 2027), 24085242(August 26, 2027), 24105106(October 22, 2027), 24066655(June 24, 2027), 24015475(January 20, 2027); 70001B-07T/10885403240997/24109019(October 2, 2027), 24109018(October 2, 2027), 24089215(August 9, 2027), 24089214(August 9, 2027), 24019266(January 16, 2027), 23109293(October 23, 2026), 24099179(September 5, 2027), 24049266(April 15, 2027), 24039248(March 14, 2027); 20350ET/10885403229756/24129026(November 13, 2027), 24079037(July 2, 2027), 24059361(May 15, 2027), 25029088(February 4, 2028), 24049028(April 2, 2027); 40000-07T/10885403225628/24119116(November 6, 2027), 24089019(August 1, 2027), 24079173(July 3, 2027), 24059480(May 28, 2027), 24059152(May 7, 2027), 24019267(January 16, 2027), 23119354(November 29, 2026), 24049267(April 15, 2027), 24039212(March 12, 2027), 23109275(October 19, 2026), 23079215(July 20, 2026); 10013364T/10885403198915/24109021(October 2, 2027), 24109020(October 2, 2027), 24099172(September 5, 2027), 24089059(August 5, 2027), 24089058(August 4, 2027), 24079175(July 3, 2027), 24079174(July 3, 2027), 24059506(May 30, 2027), 24059469(May 28, 2027), 24059151(May 7, 2027), 24039059(March 4, 2027), 24019435(January 31, 2027), 24019268(January 16, 2027), 24019433(January 30, 2027), 24019269(January 16, 2027), 23119087(November 6, 2026), 23129229(December 13, 2026), 23129230(December 13, 2026), 23059178(May 24, 2026), 23059177(May 23, 2026), 23059156(May 16, 2026), 23049061(April 26, 2026), 24059150(May 7, 2027), 24049265(April 15, 2027), 23109175(October 13, 2026), 23109006(October 2, 2026), 23099219(September 12, 2026), 23099217(September 11, 2026), 23059179(May 29, 2026); 4030B-07T/10885403240867/24089216(August 9, 2027), 24059484(May 28, 2027), 24039060(March 4, 2027), 25029393(February 27, 2028), 23109296(October 23, 2026); 22005E-07T/10885403250781/23029335(February 24, 2026)

Distribution pattern

Worldwide - US Nationwide distribution including in the states of MA, CA, IL, AK, AL, MO, FL, VA, ME, NC, MD, IA, WI, AZ, CO, KY, IN, TX, AR, NJ, NY, MT, NE, TN, PA, GA, WA, KS, MN, ID, SC, MI, SD, OH, WV, LA, ND, VT, GU, NV, UT, OK, OR, DE, NM, MS, WY and the countries of CA, AU, NZ, CN.

device · product 2 of 5

BD Texium Needle-Free Syringe: 3 mL, REF: MY8003; 5 mL, REF: MY8005; 10 mL, REF: MY8010; 20 mL, REF: MY8020; 30 mL, REF: MY8030; 60 mL, REF: MY8060

Z-1971-2025
Recall number
Z-1971-2025
Initiated
May 12, 2025
Classification
Class II
Status
Ongoing
Recalling firm
BD SWITZERLAND SARL
Quantity
1,265,400

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Sterile, single use closed system drug transfer devices, standalone or on infusion sets or bonded syringes, may leak at the interface between the closed male luer component and the needle-free connector, which may cause procedure delay, treatment interruptions, exposure to hazardous drugs, and/or underdosing of medication.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Sterile, single use closed system drug transfer devices, standalone or on infusion sets or bonded syringes, may leak at the interface between the closed male luer component and the needle-free connector, which may cause procedure delay, treatment interruptions, exposure to hazardous drugs, and/or underdosing of medication.

Code information

UDI-DI/Lot(Expiration): MY8003/10885403221651/23125144(December 5, 2026), 24015468(January 20, 2027), 24025611(February 5, 2027), 24026390(February 27, 2027), 24036107(March 25, 2027), 24045647(April 24, 2027), 24055859(May 27, 2027), 24095513(September 24, 2027), 92217803(June 27, 2025), 92225801(September 15, 2025), 92231303(November 9, 2025), 92300201(January 2, 2026), 92308702(March 28, 2026), 92309501(April 5, 2026), 92313601(May 16, 2026), 92315701(June 6, 2026) MY8005/10885403221668/23125145(December 5, 2026), 24015467(January 20, 2027), 24026387(February 27, 2027), 24045646(April 24, 2027), 24055858(May 27, 2027), 24075164(July 26, 2027), 24095512(September 24, 2027), 92215201(June 1, 2025), 92220802(July 27, 2025), 92225502(September 12, 2025), 92315102(May 31, 2026); MY8010/10885403221675/24015312(January 15, 2027), 24025610(February 5, 2027), 24045645(April 24, 2027), 24055857(May 27, 2027), 24066953(June 25, 2027), 24075165(July 26, 2027), 24085486(August 26, 2027), 24095511(September 24, 2027), 24105167(October 25, 2027), 92218601(July 5, 2025), 92222902(August 17, 2025), 92223601(August 24, 2025), 92229201(October 19, 2025), 92230701(November 3, 2025), 92231301(November 9, 2025), 92231302(November 9, 2025), 92307402(March 15, 2026), 92311601(April 26, 2026), 92315101(May 31, 2026); MY8020/10885403221682/23115069(November 3, 2026), 24015311(January 15, 2027), 24025935(February 7, 2027), 24035566(March 5, 2027), 24036106(March 25, 2027), 24045644(April 24, 2027), 24055856(May 27, 2027), 24066952(June 25, 2027), 24075163(July 26, 2027), 24085485(August 26, 2027), 24095083(September 9, 2027), 24105196(October 30, 2027), 92216401(June 13, 2025), 92217801(June 27, 2025), 92218001(June 29, 2025), 92219202(July 11, 2025), 92231901(November 15, 2025), 92300205(January 2, 2026), 92300902(January 9, 2026), 92309503(April 5, 2026), 92310101(April 11, 2026), 92313002(May 10, 2026), 92313602(May 16, 2026) MY8030/10885403230523/23115070(November 3, 2026), 23125143(December 5, 2026), 24015264(January 12, 2027), 24015644(January 27, 2027), 24026389(February 27, 2027), 24036105(March 25, 2027), 24045643(April 24, 2027), 24055855(May 27, 2027), 24075162(July 26, 2027), 24085484(August 26, 2027), 24095509(September 24, 2027), 24115110(November 5, 2027), 92225501(September 12, 2025), 92233602(December 2, 2025), 92307403(March 15, 2026), 92308701(March 28, 2026), 92309502(April 5, 2026), 92315702(June 6, 2026), 92316402(June 13, 2026), 24066951(June 25, 2027); MY8060/10885403221699/23115496(November 28, 2026), 23125393(December 18, 2026), 24015102(January 8, 2027), 24026388(February 27, 2027), 24036104(March 25, 2027), 24045366(April 23, 2027), 24055854(May 27, 2027), 24075166(July 26, 2027), 24095596(September 24, 2027), 24115111(November 14, 2027), 92216402(June 13, 2025), 92222002(August 8, 2025), 92222901(August 17, 2025), 92224303(August 31, 2025), 92224301(August 31, 2025), 92226502(September 22, 2025), 92228501(October 12, 2025), 92230702(November 3, 2025), 92230703(November 3, 2025), 92230704(November 3, 2025), 92300202(January 2, 2026), 92300203(January 2, 2026), 92306703(March 8, 2026), 92306701(March 8, 2026), 92306702(March 8, 2026), 92311602(April 26, 2026), 92312302(May 3, 2026), 92313001(May 10, 2026), 92316601(June 15, 2026), 92316602(June 15, 2026)

Distribution pattern

Worldwide - US Nationwide distribution including in the states of MA, CA, IL, AK, AL, MO, FL, VA, ME, NC, MD, IA, WI, AZ, CO, KY, IN, TX, AR, NJ, NY, MT, NE, TN, PA, GA, WA, KS, MN, ID, SC, MI, SD, OH, WV, LA, ND, VT, GU, NV, UT, OK, OR, DE, NM, MS, WY and the countries of CA, AU, NZ, CN.

device · product 3 of 5

BD Texium: 3 mL, REF: MY8003-0006; 5 mL, REF: MY8005-0006; 10 mL, REF: MY8010-0006; 20 mL, REF: MY8020-0006; 30 mL, REF: MY8030-0006; 60 mL, REF: MY8060-0006

Z-1972-2025
Recall number
Z-1972-2025
Initiated
May 12, 2025
Classification
Class II
Status
Ongoing
Recalling firm
BD SWITZERLAND SARL
Quantity
812,165

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Sterile, single use closed system drug transfer devices, standalone or on infusion sets or bonded syringes, may leak at the interface between the closed male luer component and the needle-free connector, which may cause procedure delay, treatment interruptions, exposure to hazardous drugs, and/or underdosing of medication.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Sterile, single use closed system drug transfer devices, standalone or on infusion sets or bonded syringes, may leak at the interface between the closed male luer component and the needle-free connector, which may cause procedure delay, treatment interruptions, exposure to hazardous drugs, and/or underdosing of medication.

Code information

REF/UDI-DI/Lot(Expiration): MY8003-0006/7613203018821/24015489(January 22, 2027), 24025879(February 7, 2027), 24045651(April 24, 2027), 24055852(May 27, 2027), 24066957(June 25, 2027), 24075170(July 26, 2027), 24095519(September 24, 2027), 92215801(June 7, 2025), 92217802(June 27, 2025), 92220702(July 26, 2025), 92221502(August 3, 2025), 92303201(February 1, 2026), 92303202(February 1, 2026), 92312301(May 3, 2026), 92316401(June 13, 2026); MY8005-0006/7613203018838/24025878(February 7, 2027), 24026382(February 27, 2027), 24036112(March 25, 2027), 24045650(April 24, 2027), 24055851(May 27, 2027), 24066956(June 25, 2027), 24075169(July 26, 2027), 24095517(September 24, 2027), 92229203(October 19, 2025), 92229204(October 19, 2025), 92229701(October 24, 2025), 92230602(November 2, 2025), 92304101(February 10, 2026), 92308703(March 28, 2026), 92314401(May 24, 2026), 92314402(May 24, 2026); MY8010-0006/7613203018845/24025880(February 7, 2027), 24036111(March 25, 2027), 24045652(April 24, 2027), 24066958(June 25, 2027), 24075171(July 26, 2027), 24085488(August 26, 2027), 24095518(September 24, 2027), 24105168(October 25, 2027), 92217101(June 20, 2025), 92222903(August 17, 2025), 92233202(November 28, 2025), 92233201(November 28, 2025), 92300204(January 2, 2026), 92300901(January 9, 2026), 92301901(January 19, 2026), 92301902(January 19, 2026), 92303203(February 1, 2026), 92314403(May 24, 2026); MY8020-0006/7613203018852/24025882(February 7, 2027), 24025883(February 7, 2027), 24026385(February 27, 2027), 24036110(March 25, 2027), 24045649(April 24, 2027), 24055853(May 27, 2027), 24066954(June 25, 2027), 24075168(July 26, 2027), 24085487(August 26, 2027), 24095515(September 24, 2027), 92221501(August 3, 2025), 92221503(August 3, 2025), 92222904(August 17, 2025), 92228502(October 12, 2025), 92228701(October 14, 2025), 92228702(October 14, 2025), 92229202(October 19, 2025), 92234001(December 6, 2025), 92301701(January 17, 2026), 92301702(January 17, 2026), 92302401(January 24, 2026), 92307401(March 15, 2026), 92310902(April 19, 2026); MY8030-0006/7613203026093/24025881(February 7, 2027), 24026384(February 27, 2027), 24036109(March 25, 2027), 24045648(April 24, 2027), 24066955(June 25, 2027), 24075167(July 26, 2027), 92215802(June 7, 2025), 92219201(July 11, 2025), 92220701(July 26, 2025), 92220801(July 27, 2025), 92225802(September 15, 2025), 92226501(September 22, 2025), 92231902(November 15, 2025); MY8060-0006/7613203018869/23125068(December 2, 2026), 24026383(February 27, 2027), 24045653(April 24, 2027), 24066879(June 24, 2027), 24075001(July 1, 2027), 24085489(August 26, 2027), 24095516(September 24, 2027), 24115310(November 5, 2027), 92218602(July 5, 2025), 92222001(August 8, 2025), 92223602(August 24, 2025), 92224302(August 31, 2025), 92225803(September 15, 2025), 92227002(September 27, 2025), 92227001(September 27, 2025), 92227901(October 6, 2025), 92233601(December 2, 2025), 92310102(April 11, 2026), 92310901(April 19, 2026);

Distribution pattern

Worldwide - US Nationwide distribution including in the states of MA, CA, IL, AK, AL, MO, FL, VA, ME, NC, MD, IA, WI, AZ, CO, KY, IN, TX, AR, NJ, NY, MT, NE, TN, PA, GA, WA, KS, MN, ID, SC, MI, SD, OH, WV, LA, ND, VT, GU, NV, UT, OK, OR, DE, NM, MS, WY and the countries of CA, AU, NZ, CN.

device · product 4 of 5

BD TEXIUM CLOSED MALE LUER WITH FEMALE CAP, REF: 10012241-0500

Z-1973-2025
Recall number
Z-1973-2025
Initiated
May 12, 2025
Classification
Class II
Status
Ongoing
Recalling firm
BD SWITZERLAND SARL
Quantity
1,655,300

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Sterile, single use closed system drug transfer devices, standalone or on infusion sets or bonded syringes, may leak at the interface between the closed male luer component and the needle-free connector, which may cause procedure delay, treatment interruptions, exposure to hazardous drugs, and/or underdosing of medication.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Sterile, single use closed system drug transfer devices, standalone or on infusion sets or bonded syringes, may leak at the interface between the closed male luer component and the needle-free connector, which may cause procedure delay, treatment interruptions, exposure to hazardous drugs, and/or underdosing of medication.

Code information

REF/UDI-DI/Lot(Expiration): 10012241-0500/10885403221705/23045145(April 3, 2026), 23045146(April 3, 2026), 23055091(May 4, 2026), 23065090(June 3, 2026), 23075107(July 5, 2026), 23095124(September 8, 2026), 23105027(October 3, 2026), 23115047(November 2, 2026), 23125146(December 5, 2026), 24015245(January 11, 2027), 24025887(February 7, 2027), 24026421(February 27, 2027), 24036126(March 25, 2027), 24045690(April 24, 2027), 24066893(June 25, 2027), 24075111(July 26, 2027), 24095556(September 24, 2027), 24125014(December 2, 2027), 23035049(March 1, 2026)

Distribution pattern

Worldwide - US Nationwide distribution including in the states of MA, CA, IL, AK, AL, MO, FL, VA, ME, NC, MD, IA, WI, AZ, CO, KY, IN, TX, AR, NJ, NY, MT, NE, TN, PA, GA, WA, KS, MN, ID, SC, MI, SD, OH, WV, LA, ND, VT, GU, NV, UT, OK, OR, DE, NM, MS, WY and the countries of CA, AU, NZ, CN.

device · product 5 of 5

BD TEXIUM CLOSED MALE LUER WITH FEMALE CAP, REF: 10012241; BD TEXIUM (China), REF: 10012241 CHINA

Z-1974-2025
Recall number
Z-1974-2025
Initiated
May 12, 2025
Classification
Class II
Status
Ongoing
Recalling firm
BD SWITZERLAND SARL
Quantity
1,006,512

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Sterile, single use closed system drug transfer devices, standalone or on infusion sets or bonded syringes, may leak at the interface between the closed male luer component and the needle-free connector, which may cause procedure delay, treatment interruptions, exposure to hazardous drugs, and/or underdosing of medication.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Sterile, single use closed system drug transfer devices, standalone or on infusion sets or bonded syringes, may leak at the interface between the closed male luer component and the needle-free connector, which may cause procedure delay, treatment interruptions, exposure to hazardous drugs, and/or underdosing of medication.

Code information

REF/UDI-DI/Lot(Expiration): 10012241/07613203010450/23015679(January 31, 2026), 23035092(March 6, 2026), 23055076(May 4, 2026), 23075108(July 5, 2026), 23095125(September 8, 2026), 23105028(October 3, 2026), 23125147(December 5, 2026), 24015246(January 11, 2027), 24026422(February 27, 2027), 24036125(March 25, 2027), 24045689(April 24, 2027), 24066894(June 25, 2027), 24075112(July 26, 2027), 24095557(September 24, 2027), 24115004(November 1, 2027); 10012241 CHINA/10885403238383/23095707(September 14, 2026), 24025890(February 7, 2027), 24105002(October 3, 2027)

Distribution pattern

Worldwide - US Nationwide distribution including in the states of MA, CA, IL, AK, AL, MO, FL, VA, ME, NC, MD, IA, WI, AZ, CO, KY, IN, TX, AR, NJ, NY, MT, NE, TN, PA, GA, WA, KS, MN, ID, SC, MI, SD, OH, WV, LA, ND, VT, GU, NV, UT, OK, OR, DE, NM, MS, WY and the countries of CA, AU, NZ, CN.