Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 96746

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
April 23, 2025
Product types
Device
Classifications
Class III
Statuses
Ongoing
Recalling firm wording
Beckman Coulter Ireland, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Apo B Reagent, REF: OSR6143

Z-1815-2025
Recall number
Z-1815-2025
Initiated
April 23, 2025
Classification
Class III
Status
Ongoing
Quantity
2747 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The product Value Assignment Sheet (VAS) is used in calibration of the device. An erroneous value was included in the VAS which has the potential to generate an incorrect APO B patient result which could exceed the total allowable error for the APO B assay (11.6%). The error results in results with a positive bias up to 42%. However, based on the mitigating factors the risk to health is considered to be negligible or limited.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Error in labeling

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The product Value Assignment Sheet (VAS) is used in calibration of the device. An erroneous value was included in the VAS which has the potential to generate an incorrect APO B patient result which could exceed the total allowable error for the APO B assay (11.6%). The error results in results with a positive bias up to 42%. However, based on the mitigating factors the risk to health is considered to be negligible or limited.

Code information

Lot # 2631/UDI: 15099590010409

Distribution pattern

Worldwide - US Nationwide distribution in the states of AZ, CA, CO, FL, GA, IL, IN, MD, MI, NJ, NY, OH, OK, OR, PA, TX, VA, WA, WV and the countries of Argentina, Australia, Belarus, Belgium, Bulgaria, Canada, China, Czechia, Egypt, Estonia, France, Germany, Ghana, Greece, Hong Kong, India, Ireland, Israel, Italy, Korea, Republic of, Malaysia, Netherlands, New Zealand, Russian Federation, Saudi Arabia, Slovakia, Spain, Switzerland, Taiwan, Province of China, United Kingdom of Great Britain and Northern Ireland, United States of America, Viet Nam.