Timeline bucket
May 02, 2025
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
ICU Medical, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

LifeShield Drug Library Management (DLM), LifeShield Infusion Safety Software SUite v2.1 and v2.2

Z-1763-2025
Recall number
Z-1763-2025
Initiated
May 02, 2025
Classification
Class II
Status
Ongoing
Recalling firm
ICU Medical, Inc.
Quantity
12 US, 3 OUS

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Firm has identified Drug Library Management defects in the software: 1) DLM software does not allow user to create percentage dose rate or rate change values with certain limits, potentially resulting in over-delivery to patient. 2) An unauthorized user may modify and approve a drug library, potentially leading to incorrect program parameters being used for therapy.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software Design Change

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Firm has identified Drug Library Management defects in the software: 1) DLM software does not allow user to create percentage dose rate or rate change values with certain limits, potentially resulting in over-delivery to patient. 2) An unauthorized user may modify and approve a drug library, potentially leading to incorrect program parameters being used for therapy.

Code information

LifeShield Infusion Safety Software Suite v2.1 and v2.2 UDI-DI 00887709122802

Distribution pattern