Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 96925

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
May 21, 2025
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
BIGFOOT BIOMEDICAL

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Bigfoot Unity Diabetes Management System SW-300689-01 Bigfoot Inject v1 Pen Cap LCAP Firmware 1.3.0 Release Configuration, Rev H SW-300689-02 Bigfoot Inject v1 Pen Cap RCAP Firmware 1.3.0 Release Configuration, Rev H The Bigfoot Unity Diabetes Management System that include the Pen Caps (Rapid-Acting Insulin Pen Cap [RCAP] and Long-Acting Insulin Pen Cap [LCAP]) integrates continuous monitoring with insulin dose recommendations. The Pen Caps (Rapid-Acting Insulin Pen Cap [RCAP] and Long-Acting Insulin Pen Cap [LCAP]) are part of the Bigfoot Unity Diabetes Management System.

Z-1997-2025
Recall number
Z-1997-2025
Initiated
May 21, 2025
Classification
Class II
Status
Ongoing
Recalling firm
BIGFOOT BIOMEDICAL
Quantity
57 pen caps

App-derived interpretation

Device software or design reason.device_software_design · v1.0.0
software/firmware issue

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Due to a software/firmware issue diabetes management system may provide a device error and cause the insulin pen to become unresponsive and unable to function. This error may pose a potential health risk for people living with diabetes as they may not have all information necessary to make a timely treatment decision.

Code information

Product Description / Item Number / UDI-DI code / Serial Number: Humalog/lispro/Lyumjev Bigfoot Unity OMS Cap FG-300171 00850003506418 KR242250013 Lantus-Glargine Bigfoot Unity OMS Cap FG-300176 00850003506463 SL240720021 Novolog/Aspart Bigfoot Unity OMS Cap FG-300174 00850003506449 FR242280004 Novolog/Aspart Bigfoot Unity OMS Cap FG-300174 00850003506449 FR242280005 Fiasp Bigfoot Unity DMS Cap FG-300173 00850003506432 TR240950015 Lantus-Glargine Bigfoot Unity OMS Cap FG-300176 00850003506463 SL240720022 Levmr FlxT/Tresiba-Degl Bigfoot Unity OMS Cap FG-300179 00850003506494 TL242340007 Levmr FlxT/Tresiba-Degl Bigfoot Unity OMS Cap FG-300179 00850003506494 TL242340008 Humalog/lispro/Lyumjev Bigfoot Unity OMS Cap FG-300171 00850003506418 KR242250014 Novolog/Aspart Bigfoot Unity OMS Cap FG-300174 00850003506449 FR242280006 Levmr FlxT/Tresiba-Degl Bigfoot Unity OMS Cap FG-300179 00850003506494 TL242340010 Humalog/Lispro/Lyumjev Bigfoot Unity OMS Cap FG-300171 00850003506418 KR242250015 Lantus-Glargine Bigfoot Unity OMS Cap FG-300176 00850003506463 SL240720023 Humalog/Lispro/Lyumjev Bigfoot Unity OMS Cap FG-300171 00850003506418 KR242250016 Novolog/Aspart Bigfoot Unity OMS Cap FG-300174 00850003506449 FR242280007 Humalog/Lispro/Lyumjev Bigfoot Unity OMS Cap FG-300171 00850003506418 KR242250017 Basaglar Bigfoot Unity OMS Cap FG-300175 00850003506456 FR242280009 Humalog/Lispro/Lyumjev Bigfoot Unity OMS Cap FG-300171 00850003506418 KR242250018 Humalog/Lispro/Lyumjev Bigfoot Unity OMS Cap FG-300171 00850003506418 KR242250020 Levmr FlxT/Tresiba-Degl Bigfoot Unity OMS Cap FG-300179 00850003506494 TL242340009 Novolog/Aspart Bigfoot Unity OMS Cap FG-300174 00850003506449 FR242280008 Humalog/Lispro/Lyumjev Bigfoot Unity OMS Cap FG-300171 00850003506418 KR242250021 Novolog/Aspart Bigfoot Unity OMS Cap FG-300174 00850003506449 KL241710001 Fiasp Bigfoot Unity OMS Cap FG-300173 00850003506432 TR241570002 Bigfoot Unity Cap for Novolog ASM-10046* 00850003506173 FR242280014 Bigfoot Unity Cap for Toujeo ASM-10053* 00850003506197 ST240960003 Novolog/Aspart Bigfoot Unity OMS Cap FG-300174 00850003506449 FR242280010 Levmr FlxT/Tresiba-Degl Bigfoot Unity OMS Cap FG-300179 00850003506494 TL242340011 Novolog/Aspart Bigfoot Unity OMS Cap FG-300174 00850003506449 FR242280011 Novolog/Aspart Bigfoot Unity OMS Cap FG-300174 00850003506449 FR242280012 Humalog/Lispro/Lyumjev Bigfoot Unity OMS Cap FG-300171 00850003506418 KR242250022 Humalog/Lispro/Lyumjev Bigfoot Unity OMS Cap FG-300171 00850003506418 KR242250023 Humalog/Lispro/Lyumjev Bigfoot Unity OMS Cap FG-300171 00850003506418 KR242250026 Humalog/Lispro/Lyumjev Bigfoot Unity OMS Cap FG-300171 00850003506418 KR242250024 Humalog/Lispro/Lyumjev Bigfoot Unity OMS Cap FG-300171 00850003506418 KR242250025 Novolog/Aspart Bigfoot Unity OMS Cap FG-300174 00850003506449 SL240720024 Lantus-Glargine Bigfoot Unity OMS Cap FG-300176 00850003506463 FR242280013 Novolog/Aspart Bigfoot Unity OMS Cap FG-300174 00850003506449 FR242280018 Novolog/Aspart Bigfoot Unity OMS Cap FG-300174 00850003506449 FR242280015 Humalog/Lispro/Lyumjev Bigfoot Unity OMS Cap FG-300171 00850003506418 KR242250029 Novolog/Aspart Bigfoot Unity OMS Cap FG-300174 00850003506449 FR242280016 Fiasp Bigfoot Unity DMS Cap FG-300173 00850003506432 TR241570003 Humalog/Lispro/Lyumjev Bigfoot Unity DMS Cap FG-300171 00850003506418 KR242250030 Humalog/Lispro/Lyumjev Bigfoot Unity OMS Cap FG-300171 00850003506418 KR242250027 Humalog/Lispro/Lyumjev Bigfoot Unity OMS Cap FG-300171 00850003506418 KR242250028 Levmr FlxT/Tresiba-Degl Bigfoot Unity OMS Cap FG-300179 00850003506494 TL242340012 Lantus-Glargine Bigfoot Unity OMS Cap FG-300176 00850003506463 SL240720026 Fiasp Bigfoot Unity DMS Cap FG-300173 00850003506432 TR241570004 Lantus-Glargine Bigfoot Unity DMS Cap FG-300176 00850003506463 SL240720027 Novolog/Aspart Bigfoot Unity DMS Cap FG-300174 00850003506449 FR242280017 Novolog/Aspart Bigfoot Unity DMS Cap FG-300174 00850003506449 FR242280019 Fiasp Bigfoot Unity DMS Cap FG-300173 00850003506432 TR241570005 Humalog/Lispro/Lyumjev Bigfoot Unity DMS Cap FG-300171 00850003506418 KR242250031 Humalog/Lispro/Lyumjev Bigfoot Unity DMS Cap FG-300171 00850003506418 KR242250032 Humalog/Lispro/Lyumjev Bigfoot Unity DMS Cap FG-300171 00850003506418 KR242250033 Humalog/Lispro/Lyumjev Bigfoot Unity DMS Cap FG-300171 00850003506418 KR242250034 Lantus-Glargine Bigfoot Unity DMS Cap FG-300176 00850003506463 SL240720028

Distribution pattern

U.S. Nationwide distribution in the states of AL, FL, GA, IA, IL, NY, PA, TX, and WA.