device · product 1 of 1
Pentax Medical Video Processor; Model Number: EPK-i8020c;
- Recall number
- Z-2329-2025
- Initiated
- July 16, 2025
- Classification
- Class II
- Status
- Ongoing
- Recalling firm
- Pentax of America Inc
- Quantity
- 144 units
App-derived interpretation
During endoscopic procedures using a combination of the video processor EPK-i8020c and i20c series video endoscope, the observed image can become reddish or dark. Users have observed smoke like steam and noted that the light guide at the tip is hot during use / after the endoscope is removed from the patient and may cause thermal injury to the patient s mucous membrane.
Official device-enrichment evidence · Sourced
Device Design
Inspect official wording and provenance
Reason for recall
During endoscopic procedures using a combination of the video processor EPK-i8020c and i20c series video endoscope, the observed image can become reddish or dark. Users have observed smoke like steam and noted that the light guide at the tip is hot during use / after the endoscope is removed from the patient and may cause thermal injury to the patient s mucous membrane.
Code information
Model Number: EPK-i8020c; UDI-DI: 04961333247974; Serial Numbers: B0023Z1657, D0023Z0245, D0023Z0206, D0023Z0125, D0023Z0123, D0023Z0119, D0023Z0118, D0023Z0117, D0023Z0114, D0023Z0112, D0023Z0109, D0023Z0108, D0023Z0011, C0023Z0087, C0023Z0085, C0023Z0074, C0023Z0072, C0023Z0021, B0023Z1712, B0023Z1711, B0023Z1670, B0023Z1665, B0023Z1659, B0023Z1653, B0023Z1628, b0023z1501, B0023Z1195, B0023Z1707, C0023Z0084, B0023Z1732, B0023Z1669, D0023Z0137, D0023Z0141, D0158Z0049, D0158Z0033, E0023Z0022, D0158Z0053, D0158Z0019, E0023Z0026, E0023Z0008, D0023Z0248, D0023Z0241, D0023Z0222, D0023Z0155, B0023Z1276, C0023Z0082, D0023Z0233, B0023Z1461, D0023Z0139, D0023Z0140, D0023Z0122, B0023Z1709, E0023Z0020, E0023Z0009, E0023Z0002, D0023Z0142, E0023Z0024, E0023Z0018, E0023Z0013, E0023Z0004, C0023Z0095, C0023Z0078, E0023Z0007, E0158Z0057, E0158Z0035, B0023Z1735, B0023Z1717, B0023Z1648, B0023Z1639, B0023Z1631, C0158Z0047, C0158Z0042, B0023Z1481, B0023Z1478, B0023Z1451, B0023Z1335, B0023Z1193, B0023Z1288, B0023Z1258, B0023Z1296, B0023Z1290, B0023Z1286, B0023Z1280, B0023Z1279, B0023Z1256, B0023Z1250, B0023Z1245, B0023Z1676, B0023Z1573, B0023Z1565, B0023Z1656, B0023Z1694, D0158Z0055, B0023Z1577, B0023Z1570, B0023Z1682, B0023Z1678, B0023Z1661, B0023Z1637, D0158Z0048, D0158Z0047, E0023Z0016, E0023Z0001, D0023Z0071, D0023Z0150, D0023Z0148, D0023Z0143, B0023Z1277, D0023Z0152, D0023Z0144, D0023Z0106, D0023Z0102, B0023Z1667, B0023Z1652, B0023Z1645, B0023Z1727, B0023Z1723, B0023Z1702, E0023Z0003, D0023Z0244, D0023Z0232, C0023Z0091, D0023Z0208, B0023Z1655, D0023Z0219, D0023Z0017, D0023Z0008, C0023Z0098, C0023Z0097, C0023Z0096, C0023Z0094, C0023Z0092, B0023Z1716, B0023Z1698, B0023Z1697, B0023Z1696, B0023Z1671, B0023Z1255, B0023Z1247, B0023Z1291, B0023Z1289, B0023Z1287, B0023Z1285, B0023Z1243;
Distribution pattern
US Distribution to states of: AZ, CA, DC, FL IA, ID, IL, KS, MA, MD, MI, MO, NC, NJ, NM, NY, OK, PA, TX, WA.