device · product 1 of 1
HemosIL LMW Heparin Controls; Part Number: 0020300200;
- Recall number
- Z-2323-2025
- Initiated
- February 03, 2025
- Classification
- Class II
- Status
- Ongoing
- Recalling firm
- Instrumentation Laboratory
- Quantity
- 41,824 units
App-derived interpretation
Multiple complaints indicating lower than expected quality control (QC) results and in some instances QC recovery below the lower limit of the package insert acceptance range when testing is performed with HemosIL Liquid Anti-Xa and HemosIL Heparin Calibrator. In some complaints, failure to pass QC resulted in patient testing delays.
Official device-enrichment evidence · Sourced
Process control
Inspect official wording and provenance
Reason for recall
Multiple complaints indicating lower than expected quality control (QC) results and in some instances QC recovery below the lower limit of the package insert acceptance range when testing is performed with HemosIL Liquid Anti-Xa and HemosIL Heparin Calibrator. In some complaints, failure to pass QC resulted in patient testing delays.
Code information
Part Number: 0020300200; UDI-DI: 08426950472490; Lot Numbers: N0330228, N0632773, N0935135, N1036592, N0148986, N0441476, N0643673, N0845161, N1147397;
Distribution pattern
Worldwide - US Nationwide and the countries of Algeria, Andorra, Argentina, Australia, Austria, Belarus, Brazil, Brunei, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Denmark, Finland, Germany, Gibraltar, Greece, Guatemala, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Korea, Kuwait, Kyrgyzstan, Lebanon, Lithuania, Macao, Malaysia, Mexico, Panama, Paraguay, Peru, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, South Africa, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, United Kingdom, Uruguay, Vietnam.