Event 97187

22 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
June 20, 2025
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
GE Medical Systems, LLC

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

22 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 22

GE Signa Artist, Nuclear Magnetic Resonance Imaging System

Z-2299-2025
Recall number
Z-2299-2025
Initiated
June 20, 2025
Classification
Class II
Status
Ongoing
Recalling firm
GE Medical Systems, LLC
Quantity
760 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Equipment maintenance

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

Code information

GTIN: 00840682123129, 00840682146104, 00195278117021, 00840682123457, 00195278126443, 00195278210036, Not available, and Not applicable; All System ID numbers that GE HealthCare performed planned maintenance on since March 1, 2023

Distribution pattern

Worldwide distribution.

device · product 2 of 22

GE Signa Voyager, Nuclear Magnetic Resonance Imaging System

Z-2300-2025
Recall number
Z-2300-2025
Initiated
June 20, 2025
Classification
Class II
Status
Ongoing
Recalling firm
GE Medical Systems, LLC
Quantity
280 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Equipment maintenance

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

Code information

GTIN: 00840682108607, 00195278124609, 00195278248923, 00840682118118, Not applicable, Not available; All System ID numbers that GE HealthCare performed planned maintenance on since March 1, 2023

Distribution pattern

Worldwide distribution.

device · product 3 of 22

GE Discovery MR450 1.5T, Nuclear Magnetic Resonance Imaging System

Z-2301-2025
Recall number
Z-2301-2025
Initiated
June 20, 2025
Classification
Class II
Status
Ongoing
Recalling firm
GE Medical Systems, LLC
Quantity
19 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Equipment maintenance

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

Code information

GTIN: Not available, Not applicable; All System ID numbers that GE HealthCare performed planned maintenance on since March 1, 2023

Distribution pattern

Worldwide distribution.

device · product 4 of 22

GE Optima MR450w 1.5T, Nuclear Magnetic Resonance Imaging System

Z-2302-2025
Recall number
Z-2302-2025
Initiated
June 20, 2025
Classification
Class II
Status
Ongoing
Recalling firm
GE Medical Systems, LLC
Quantity
1372 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Equipment maintenance

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

Code information

00195278023643, 00195278117021, 00195278126443, 00195278210036, 00195278229519, 00840682115971, 00840682118118, 00840682118132, 00840682123129, 00840682123457, 00840682123457, 00840682146104, Not applicable, Not available; All System ID numbers that GE HealthCare performed planned maintenance on since March 1, 2023

Distribution pattern

Worldwide distribution.

device · product 5 of 22

GE SIGNA Architect, Nuclear Magnetic Resonance Imaging System

Z-2303-2025
Recall number
Z-2303-2025
Initiated
June 20, 2025
Classification
Class II
Status
Ongoing
Recalling firm
GE Medical Systems, LLC
Quantity
758 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Equipment maintenance

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

Code information

GTIN: 00195278023643, 00195278117021, 00840682102728, 00840682103817, 00840682118118, 00840682122702, 00840682123129, 00840682123440, 00840682146197, 00840682147095, Not Applicable, Not available; All System ID numbers that GE HealthCare performed planned maintenance on since March 1, 2023

Distribution pattern

Worldwide distribution.

device · product 6 of 22

GE SIGNA Premier, Nuclear Magnetic Resonance Imaging System

Z-2304-2025
Recall number
Z-2304-2025
Initiated
June 20, 2025
Classification
Class II
Status
Ongoing
Recalling firm
GE Medical Systems, LLC
Quantity
214 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Equipment maintenance

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

Code information

GTIN: 00840682146999, 00195278010797, 00840682135269, 00840682146999, Not applicable, Not available; All System ID numbers that GE HealthCare performed planned maintenance on since March 1, 2023

Distribution pattern

Worldwide distribution.

device · product 7 of 22

GE SIGNA UHP, Nuclear Magnetic Resonance Imaging System

Z-2305-2025
Recall number
Z-2305-2025
Initiated
June 20, 2025
Classification
Class II
Status
Ongoing
Recalling firm
GE Medical Systems, LLC
Quantity
3 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Equipment maintenance

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

Code information

GTIN: Not Applicable; All System ID numbers that GE HealthCare performed planned maintenance on since March 1, 2023

Distribution pattern

Worldwide distribution.

device · product 8 of 22

GE Discovery MR750 3.0T, Nuclear Magnetic Resonance Imaging System

Z-2306-2025
Recall number
Z-2306-2025
Initiated
June 20, 2025
Classification
Class II
Status
Ongoing
Recalling firm
GE Medical Systems, LLC
Quantity
330 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Equipment maintenance

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

Code information

GTIN: 00195278229519, 00840682115872, 00840682115971, 00840682118187, 00840682123129, 00840682146999, Not applicable, Not available; All System ID numbers that GE HealthCare performed planned maintenance on since March 1, 2023

Distribution pattern

Worldwide distribution.

device · product 9 of 22

GE Discovery MR750w 3.0T, Nuclear Magnetic Resonance Imaging System

Z-2307-2025
Recall number
Z-2307-2025
Initiated
June 20, 2025
Classification
Class II
Status
Ongoing
Recalling firm
GE Medical Systems, LLC
Quantity
519 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Equipment maintenance

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

Code information

GTIN: 00195278023643, 00840682102704, 00840682102728, 00840682103787, 00840682103817, 00840682123440, Not applicable, Not Available; All System ID numbers that GE HealthCare performed planned maintenance on since March 1, 2023

Distribution pattern

Worldwide distribution.

device · product 10 of 22

GE SIGNA Hero, Nuclear Magnetic Resonance Imaging System

Z-2308-2025
Recall number
Z-2308-2025
Initiated
June 20, 2025
Classification
Class II
Status
Ongoing
Recalling firm
GE Medical Systems, LLC
Quantity
56 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Equipment maintenance

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

Code information

GTIN: 00195278472083, 00195278486813, 001952784868131, Not Applicable; All System ID numbers that GE HealthCare performed planned maintenance on since March 1, 2023

Distribution pattern

Worldwide distribution.

device · product 11 of 22

GE SIGNA PET/MR, Tomographic Imager Combining Emission Computed Tomography with Nuclear Magnetic Resonance

Z-2309-2025
Recall number
Z-2309-2025
Initiated
June 20, 2025
Classification
Class II
Status
Ongoing
Recalling firm
GE Medical Systems, LLC
Quantity
97 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Equipment maintenance

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

Code information

GTIN: 00195278648877, 00840682105378, 00840682123440, 00840682135283, Not applicable, Not available; All System ID numbers that GE HealthCare performed planned maintenance on since March 1, 2023

Distribution pattern

Worldwide distribution.

device · product 12 of 22

GE SIGNA Artist Evo, Nuclear Magnetic Resonance Imaging System

Z-2310-2025
Recall number
Z-2310-2025
Initiated
June 20, 2025
Classification
Class II
Status
Ongoing
Recalling firm
GE Medical Systems, LLC
Quantity
138 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Equipment maintenance

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

Code information

GTIN 00195278481382, 00840682118118, 00840682123129, 00840682147095, Not Applicable, GTIN Not Available; All System ID numbers that GE HealthCare performed planned maintenance on since March 1, 2023

Distribution pattern

Worldwide distribution.

device · product 13 of 22

GE SIGNA Voyager AIR (China Only), Nuclear Magnetic Resonance Imaging System

Z-2311-2025
Recall number
Z-2311-2025
Initiated
June 20, 2025
Classification
Class II
Status
Ongoing
Recalling firm
GE Medical Systems, LLC
Quantity
0

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Equipment maintenance

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

Code information

GTIN: unknown; All System ID numbers that GE HealthCare performed planned maintenance on since March 1, 2023

Distribution pattern

Worldwide distribution.

device · product 14 of 22

GE SIGNA Architect AIR (China Only), Nuclear Magnetic Resonance Imaging System

Z-2312-2025
Recall number
Z-2312-2025
Initiated
June 20, 2025
Classification
Class II
Status
Ongoing
Recalling firm
GE Medical Systems, LLC
Quantity
151 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Equipment maintenance

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

Code information

GTIN: 00840682102728, 00840682103787, Not available; All System ID numbers that GE HealthCare performed planned maintenance on since March 1, 2023

Distribution pattern

Worldwide distribution.

device · product 15 of 22

GE SIGNA Premier XT (China Only), Nuclear Magnetic Resonance Imaging System

Z-2313-2025
Recall number
Z-2313-2025
Initiated
June 20, 2025
Classification
Class II
Status
Ongoing
Recalling firm
GE Medical Systems, LLC
Quantity
2 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Equipment maintenance

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

Code information

GTIN: Not available; All System ID numbers that GE HealthCare performed planned maintenance on since March 1, 2023

Distribution pattern

Worldwide distribution.

device · product 16 of 22

GE SIGNA Premier Elite (China Only), Nuclear Magnetic Resonance Imaging System

Z-2314-2025
Recall number
Z-2314-2025
Initiated
June 20, 2025
Classification
Class II
Status
Ongoing
Recalling firm
GE Medical Systems, LLC
Quantity
0

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Equipment maintenance

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

Code information

GTIN: unknown; All System ID numbers that GE HealthCare performed planned maintenance on since March 1, 2023

Distribution pattern

Worldwide distribution.

device · product 17 of 22

GE SIGNA Premier MAX (China Only), Nuclear Magnetic Resonance Imaging System

Z-2315-2025
Recall number
Z-2315-2025
Initiated
June 20, 2025
Classification
Class II
Status
Ongoing
Recalling firm
GE Medical Systems, LLC
Quantity
0

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Equipment maintenance

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

Code information

GTIN: Unknown; All System ID numbers that GE HealthCare performed planned maintenance on since March 1, 2023

Distribution pattern

Worldwide distribution.

device · product 18 of 22

GE SIGNA Hero XT (China Only), Nuclear Magnetic Resonance Imaging System

Z-2316-2025
Recall number
Z-2316-2025
Initiated
June 20, 2025
Classification
Class II
Status
Ongoing
Recalling firm
GE Medical Systems, LLC
Quantity
N/A

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Equipment maintenance

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

Code information

GTIN: Unknown; All System ID numbers that GE HealthCare performed planned maintenance on since March 1, 2023

Distribution pattern

Worldwide distribution.

device · product 19 of 22

GE SIGNA Hero Elite (China Only), Nuclear Magnetic Resonance Imaging System

Z-2317-2025
Recall number
Z-2317-2025
Initiated
June 20, 2025
Classification
Class II
Status
Ongoing
Recalling firm
GE Medical Systems, LLC
Quantity
0

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Equipment maintenance

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

Code information

GTIN: Unknown; All System ID numbers that GE HealthCare performed planned maintenance on since March 1, 2023

Distribution pattern

Worldwide distribution.

device · product 20 of 22

GE SIGNA Hero MAX (China Only), Nuclear Magnetic Resonance Imaging System

Z-2318-2025
Recall number
Z-2318-2025
Initiated
June 20, 2025
Classification
Class II
Status
Ongoing
Recalling firm
GE Medical Systems, LLC
Quantity
0

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Equipment maintenance

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

Code information

GTIN: Unknown; All System ID numbers that GE HealthCare performed planned maintenance on since March 1, 2023

Distribution pattern

Worldwide distribution.

device · product 21 of 22

GE SIGNA PET/MR AIR (China Only), Nuclear Magnetic Resonance Imaging System

Z-2319-2025
Recall number
Z-2319-2025
Initiated
June 20, 2025
Classification
Class II
Status
Ongoing
Recalling firm
GE Medical Systems, LLC
Quantity
0

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Equipment maintenance

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

Code information

GTIN: Unknown; All System ID numbers that GE HealthCare performed planned maintenance on since March 1, 2023

Distribution pattern

Worldwide distribution.

device · product 22 of 22

GE SIGNA PET/MR AIR Plus (China Only), Nuclear Magnetic Resonance Imaging System

Z-2320-2025
Recall number
Z-2320-2025
Initiated
June 20, 2025
Classification
Class II
Status
Ongoing
Recalling firm
GE Medical Systems, LLC
Quantity
0

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Equipment maintenance

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

Code information

GTIN: Unknown; All System ID numbers that GE HealthCare performed planned maintenance on since March 1, 2023

Distribution pattern

Worldwide distribution.