device · product 1 of 2
Brand Name: Atlantis IO FLO-S Product Name: Atlantis IO FLO-S Kit Model/Catalog Number: A04B / 68020033 Software Version: n/a Product Description: The Atlantis intraoral feature locating object for suprastructures (IO FLO-S) is intended to support the prosthetic procedure with the aim of restoring chewing function in partial or complete edentulous patients, by enabling digital impression taking on implant or abutment level for all positions in the mouth of the patient or on the master model. Component: not a component
The defect in the thread area will not allow the device to fully engage with an implant or analog (implant replica used for prosthetic modeling). The threads will only engage approximately 2 rotations as opposed to full engagement which is approximately 5 rotations. This equates to approximately 1mm difference in depth of engagement.
These labels are deterministic app interpretations, not FDA categories.
Official device-enrichment evidence · Unknown
An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.
The defect in the thread area will not allow the device to fully engage with an implant or analog (implant replica used for prosthetic modeling). The threads will only engage approximately 2 rotations as opposed to full engagement which is approximately 5 rotations. This equates to approximately 1mm difference in depth of engagement.
Worldwide - US Nationwide distribution in the states of IL, LA, MA, MN, MO, ND, NJ, NM, OH, OK, PA, TX, WV and the countries of AT, AU, BE, CA, DE, DK, ES, FI, FR, IT, JP, NL, NO, PL, QA, SE, UA, UK.
device · product 2 of 2
Brand Name: Atlantis IO FLO Product Name: Atlantis IO FLO-S Kit Model/Catalog Number: A04S / 6802034 Product Description: The Atlantis intraoral feature locating object for suprastructures (IO FLO-S) is intended to support the prosthetic procedure with the aim of restoring chewing function in partial or complete edentulous patients, by enabling digital impression taking on implant or abutment level for all positions in the mouth of the patient or on the master model. Component: not a component
The defect in the thread area will not allow the device to fully engage with an implant or analog (implant replica used for prosthetic modeling). The threads will only engage approximately 2 rotations as opposed to full engagement which is approximately 5 rotations. This equates to approximately 1mm difference in depth of engagement.
These labels are deterministic app interpretations, not FDA categories.
Official device-enrichment evidence · Unknown
An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.
The defect in the thread area will not allow the device to fully engage with an implant or analog (implant replica used for prosthetic modeling). The threads will only engage approximately 2 rotations as opposed to full engagement which is approximately 5 rotations. This equates to approximately 1mm difference in depth of engagement.
Worldwide - US Nationwide distribution in the states of IL, LA, MA, MN, MO, ND, NJ, NM, OH, OK, PA, TX, WV and the countries of AT, AU, BE, CA, DE, DK, ES, FI, FR, IT, JP, NL, NO, PL, QA, SE, UA, UK.