Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 97390

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
August 04, 2025
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Beckman Coulter Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

DxC 500 AU Module w/ISE, DxC 500i, With ISE, REF: C63522 and DxC 500 AU Module, DxC 500i, Without ISE, REF: C6352; part of the DxC 500i Clinical Analyzer

Z-0032-2026
Recall number
Z-0032-2026
Initiated
August 04, 2025
Classification
Class II
Status
Ongoing
Recalling firm
Beckman Coulter Inc.
Quantity
116

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
On a clinical analyzer, when ordering a Clinical Chemistry (CC) combination test simultaneously with an Immunoassay (IA) test, if the IA test is processed between the CC tests, CC sample status will remain In Progress with no errors, may cause processing delays.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software Design Change

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

On a clinical analyzer, when ordering a Clinical Chemistry (CC) combination test simultaneously with an Immunoassay (IA) test, if the IA test is processed between the CC tests, CC sample status will remain In Progress with no errors, may cause processing delays.

Code information

REF/UDI-DI: C63522/14987666545089, C63521/14987666545072. Software Versions: SW 1.3.3 and 1.3.4

Distribution pattern

Internationally distribution to the countries of Poland, India, Dominican Republic, Lebanon, Italy, Brazil, El Salvador, Saudi Arabia, Germany, France, Nepal, Canada, Chile, Spain, Malaysia, Switzerland, Morocco, Philippines, Curacao, Qatar, United Arab Emirates, Hungary, Czechia, Oman, New Zealand, Thailand, Indonesia, Algeria, United Kingdom of Great Britain and Northern Ireland, Namibia, Turkey, Republic of Korea