Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 97469

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
August 06, 2025
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Maquet Cardiovascular, LLC

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Vasoview Hemopro 3 Endoscopic Vessel Harvesting System Model No. C-VH-6000. The VASOVIEW HEMOPRO 3 Endoscopic Vessel Harvesting (EVH) System (VH-6000, VH-6001, and VH-6010) is a vessel harvesting kit designed for use with a 7 mm endoscope to cut and cauterize vessel branches. The HP3 Harvesting Tool features two curved, insulated jaws with heating elements that operate through heat and pressure rather than radiofrequency energy.

Z-2636-2025
Recall number
Z-2636-2025
Initiated
August 06, 2025
Classification
Class II
Status
Ongoing
Quantity
98 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Reports of out-of-box failures discovered during the limited launch phase. Failure modes included angled deformity at the attachment site of the Harvesting Tool Jaws and the distal end of the Harvesting Tool Shaft, jaws not fully closed out of box, and heater wire lifted/flexed away from jaws.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Reports of out-of-box failures discovered during the limited launch phase. Failure modes included angled deformity at the attachment site of the Harvesting Tool Jaws and the distal end of the Harvesting Tool Shaft, jaws not fully closed out of box, and heater wire lifted/flexed away from jaws.

Code information

Model No. C-VH-6000; UDI: 00607567701274; Lot/Serial No. 3000486429, 3000487231, 3000487726, 3000488639.

Distribution pattern

US Nationwide distribution in the states of FL, IL, MD & TX.