Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 97669

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
October 17, 2025
Product types
Device
Classifications
Class I
Statuses
Ongoing
Recalling firm wording
CareFusion 303, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

BD Alaris infusion Pump Module 8100, REF: 8100; KIT, BEZEL ASSY 8100BD, REF: 49000437; KIT ASSY ONE PIECE BEZEL 8100, REF: 49000270. Used with the following new and updated documents: BD Alaris Infusion System Inspection Safety Reminder, Part: BD-163124, Revision: 00; Cleaning and Disinfecting Procedure, Part: P00000864; Cleaning and Disinfecting Quick Reference Guide

Z-0430-2026
Recall number
Z-0430-2026
Initiated
October 17, 2025
Classification
Class I
Status
Ongoing
Recalling firm
CareFusion 303, Inc.
Quantity
2,574,396

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
If infusion pump is dropped or severely jarred this may damage the pump module bezel assembly, which can cause under-infusion, over-infusion, unregulated flow, or pump module failure to calibrate; so dropped/jarred pumps should be removed from use, tested, and inspected by qualified service personnel prior to reuse; tip sheet, cleaning and disinfecting procedure and reference guide updated.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

If infusion pump is dropped or severely jarred this may damage the pump module bezel assembly, which can cause under-infusion, over-infusion, unregulated flow, or pump module failure to calibrate; so dropped/jarred pumps should be removed from use, tested, and inspected by qualified service personnel prior to reuse; tip sheet, cleaning and disinfecting procedure and reference guide updated.

Code information

All Alaris Pump Modules distributed after June 19, 2004. REF/UDI-DI: 8100/10885403810015, 10885403810046, 10885403810039, 10885403222054, 10885403517723, 10885403213212, 10885403211423; 49000437/10885403471018; 49000270/10885403247743

Distribution pattern

Worldwide distribution. US nationwide including Puerto Rico, Guam, American Samoa, Northern Mariana Islands; and countries of SG, BH, AE, JP, CA, IT, QA, PH, DE, TW, PK, KR, BW, SA, MY, IL, NZ, AU, IN, GB, CN, NL, FR, ZA, BE, MX, CH, GI, KW, and TR.