device · product 1 of 1
Zenition 50; Product Number (REF): 718096; Interventional fluoroscopic x-ray system
Over time some systems may experience loss of imaging functionality and/or poor image quality due to potential corrosion of the Image Intensifier Television (IITV) control board.
These labels are deterministic app interpretations, not FDA categories.
Over time some systems may experience loss of imaging functionality and/or poor image quality due to potential corrosion of the Image Intensifier Television (IITV) control board.
Code information
Product Number (REF): 718096; UDI-DI: 00884838091535; All serial numbers;
Distribution pattern
Distribution US states of MO and TX; Belgium, Canada, Fiji, Germany, India, Malaysia, Mexico, Morocco, South Africa, South Korea, Thailand, United Kingdom, and Uzbekistan.