device · product 1 of 1
Philips Azurion 7M20 systems with FlexArm ceiling-mounted system. Model Number: 722234;
- Recall number
- Z-0719-2026
- Initiated
- October 02, 2025
- Classification
- Class II
- Status
- Ongoing
- Recalling firm
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Quantity
- 172 units
App-derived interpretation
The motorized longitudinal movement of the FlexArm stand may be inconsistent (not smooth) and eventually may become unavailable. This issue is due to grease leakage from the bearings and/or anti-corrosion oil applied on the bearings, because of which the lubrication on the ceiling rail may be excessive, reducing the friction between the rails and the friction wheel. Expansion to add additional model.
Official device-enrichment evidence · Sourced
Equipment maintenance
Inspect official wording and provenance
Reason for recall
The motorized longitudinal movement of the FlexArm stand may be inconsistent (not smooth) and eventually may become unavailable. This issue is due to grease leakage from the bearings and/or anti-corrosion oil applied on the bearings, because of which the lubrication on the ceiling rail may be excessive, reducing the friction between the rails and the friction wheel. Expansion to add additional model.
Code information
Model Number: 722234;
Distribution pattern
Domestic: AL, AR, AZ, CA, CO, FL, GA, HI, IL, IN, KS, KY, MD, MN, MO, NM, NY, OH, OK, PA, RI, TX, UT, VA, WA, WI, WV, WY; International: Australia, Austria, Belgium, Bulgaria, Canada, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, Iraq, Italy, Japan, Netherlands, Norway, Philippines, Poland, Romania, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, United Kingdom, Vietnam;