device · product 1 of 1
MAGNETOM Cima.X Upgrade. Model Number: 11689304.
- Recall number
- Z-0431-2026
- Initiated
- October 01, 2025
- Classification
- Class I
- Status
- Ongoing
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Quantity
- 4 system (1 US; 3 OUS)
App-derived interpretation
There is a potential for an ice blockage to form or currently exist within the primary and/or the auxiliary venting paths of the magnet helium containment venting system. In the event of a quench when both venting paths are blocked, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
Official device-enrichment evidence · Unknown
An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.
Inspect official wording and provenance
Reason for recall
There is a potential for an ice blockage to form or currently exist within the primary and/or the auxiliary venting paths of the magnet helium containment venting system. In the event of a quench when both venting paths are blocked, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
Code information
Model Number: 11689304. UDI: (01)04056869975245(21)237304. Serial Number: 237304.
Distribution pattern
US distribution to California. International distribution to Australia, Germany, Great Britian.