Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 97968

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
October 27, 2025
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Medline Industries, LP

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

MEDLINE PERFUSION SYRINGE, REF: DYNJ39223C

Z-0886-2026
Recall number
Z-0886-2026
Initiated
October 27, 2025
Classification
Class II
Status
Ongoing
Recalling firm
Medline Industries, LP
Quantity
15 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing BD Luer Tip Caps. BD has issued a recall of their Luer Tip Caps after confirming through internal testing that certain BD Luer Tip Cap trays failed routine biocompatibility testing. Although the tip cap covers a small surface area of the syringe and does not come in direct contact with patients, there is the potential that the failure of a biocompatibility test may have varying consequences ranging from no harm or clinical impact to potential for sensitization, irritation, pyrogenicity, acute systemic toxicity or hemolysis depending on the nature and duration of contact with the syringe.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing BD Luer Tip Caps. BD has issued a recall of their Luer Tip Caps after confirming through internal testing that certain BD Luer Tip Cap trays failed routine biocompatibility testing. Although the tip cap covers a small surface area of the syringe and does not come in direct contact with patients, there is the potential that the failure of a biocompatibility test may have varying consequences ranging from no harm or clinical impact to potential for sensitization, irritation, pyrogenicity, acute systemic toxicity or hemolysis depending on the nature and duration of contact with the syringe.

Code information

UDI/DI 10193489621532 (each) 40193489621533 (case), Lot Number 25FMJ081

Distribution pattern

US Nationwide distribution in the state of CA.