Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 98101

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
November 18, 2025
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
AGFA Healthcare Corp.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Brand name: AGFA Digital Radiography X- Ray system DR 800 with MUSICA Dynamic Common Name: DR 800

Z-0923-2026
Recall number
Z-0923-2026
Initiated
November 18, 2025
Classification
Class II
Status
Ongoing
Recalling firm
AGFA Healthcare Corp.
Quantity
35

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
It is possible to exceed the maximum allowed Air Kerma rate (AKR) of 88mGy/min for pulsed fluoro exams. This can happen in case the framerate is increased on the NX workstation when ABS is not enabled.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Radiation Control for Health and Safety Act

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

It is possible to exceed the maximum allowed Air Kerma rate (AKR) of 88mGy/min for pulsed fluoro exams. This can happen in case the framerate is increased on the NX workstation when ABS is not enabled.

Code information

N/A

Distribution pattern

U.S. Nationwide distribution.