Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 98132

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
December 08, 2025
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
IMRIS Imaging Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Brand Name: HFD100 Product Name: Head Fixation Device HFD100 Model/Catalog Number: 119695-000, 119629-000, 119630-000, 113803-000, 113803-600, 113802-000, 113802-600 Software Version: Not Applicable Product Description: The HFD100 Head Fixation Device provides rigid skeletal fixation within the optimal imaging envelope while introducing minimal artifacts in the acquired images. There are two different assemblies for the HFD100. The standard HFD100 is equipped with a table adapter for use with either the operating room table or the angiography table. For sites with a Hill-Rom TruSystem 7500 Hybrid Neuro MR Operating Table (ORT400), the HFD100 has a split block clamp assembly (Trumpf Mount Clamp Assembly) which connects directly to the table back section. The HFD100 system is also comprised of the linkage components to mount the skull clamp to the table adapter, and the 3-pin skull clamp. The linkage system is used to position the skull clamp in the necessary position, without colliding with the magnet bore. Component: Not a component.

Z-1034-2026
Recall number
Z-1034-2026
Initiated
December 08, 2025
Classification
Class II
Status
Ongoing
Recalling firm
IMRIS Imaging Inc
Quantity
122

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Torque screw may crack or separate. If used, an affected screw could lead to inadequate head stabilization, compromised surgical precision, procedural delays, requirement for alternate head fixation methods, and risk of patient injury.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Torque screw may crack or separate. If used, an affected screw could lead to inadequate head stabilization, compromised surgical precision, procedural delays, requirement for alternate head fixation methods, and risk of patient injury.

Code information

Affected UDI-DI: 00857534006011 00857534006592 00857534006608 00857534006615 Affected serial numbers: 10006206 10005998 10003999 10004026 10003796 10003811 10006629 10006630 10007514 10007515 10002486 10003048 10004215 10004216 10003794 10004231 10012064 10012310 10012311 10010990 10011142 10007914 10007971 10002391 10002969 10009132 10010058 10011570 10005134 10005711 10009569 10009878 10010057 10003903 10003906 10003546 10006187 10012838 10010639 10010666 10011925 10006692 10007246 10011463 10005092 10007237 10007509 10008416 10003407 10003793 10004195 10004975 10004993 10005136 10010381 10007906 10013179 10007510 10007511 10007245 10004449 10004961 10005849 10005851 10010747 10006730 10011187 10011307 10009593 10010003 10008073 10008354 10011546 10011571 10006207 10012127 10006631 10007777 10006411 10004466 10005445 10011645 10009015 10004291 10003985 10003987 10006204 10006205 10007235 10007236 10006412 10004775 10004323 10004942 10003936 10003937 10007130 10002695 10007126 10007128 10009228 10009210 10009583 10005480 10005481 10012839 10008645 10008670 10006693 10006694 10004317 10002403 10004673 10004671 10010572 10006212 10006203 10002968 10002715 10011186 10011265 10010059

Distribution pattern

Worldwide distribution - US Nationwide and the countries of Canada, Sweden, Switzerland, Japan, Australia, Qatar, and China.