device · product 1 of 1
Mazor X robotic guidance system REF: TPL0059
- Recall number
- Z-1122-2026
- Initiated
- December 10, 2025
- Classification
- Class II
- Status
- Ongoing
- Recalling firm
- Mazor Robotics Ltd
- Quantity
- 549 systems
App-derived interpretation
Software errors that can result in incorrect
Official device-enrichment evidence · Unknown
An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.
Inspect official wording and provenance
Reason for recall
Software errors that can result in incorrect surgical instrument positioning during spinal surgery.
Code information
Software versions: 5.0.1, 5.1.2, 5.1.3/ UDI: 07290109180465, 07290109184524, 07290109181158, 00763000635169, 07290109183213, 07290109184517, 00763000419134, 00763000431761, 07290109184098, 07290109184838, 07290115751376, 07290115751895
Distribution pattern
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Brazil, Canada, Canary Islands, China, Croatia, France, Germany, Greece, Hong Kong, Hungary, India, Israel, Italy, Japan, Malaysia, Mexico, Netherlands, Poland, Portugal, Puerto Rico, Romania, Singapore, Slovakia, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom, United States, Vietnam.