Exact FDA event identifier

Event 98232

7 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
December 11, 2025
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Fujirebio Diagnostics, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

7 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 7

Product Name: Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio Model/Catalog Number: CL0073 Software Version: Not Applicable Product Description: Lumipulse G pTau217/¿-Amyloid 1-42 Plasma Ratio is an in vitro diagnostic (IVD) test that combines the test results of the Lumipulse G pTau 217 Plasma assay and the Lumipulse G ¿-Amyloid 1-42-N Plasma assay from the same patient specimen (K2EDTA plasma sample) into a numerical ratio. The Lumipulse G pTau 217/¿-Amyloid 1-42 Plasma Ratio Instructions For Use (IFU), CL0072, contains instructions on how to manually calculate the ratio and is supplied with each individual Lumipulse G assay. Lumipulse G pTau217/¿-Amyloid 1-42 Plasma Ratio, Item# CL0073 is comprised of the following kits, which are imported: Kit Description, Item#: Imported from Fujirebio Europe (FRE): Lumipulse G pTau 217 Plasma Immunoreaction Cartridges 81550 Lumipulse G pTau 217 Plasma Calibrators set 81557 Lumipulse pTau 217 Plasma Controls 81558 Lumipulse ¿-Amyloid Plasma Controls 81559 Imported from Fujirebio Japan (FRI): Lumipulse G ¿-Amyloid 1-42-N Plasma Immunoreaction Cartridges 235447 Lumipulse G ¿-Amyloid 1-42-N Plasma Calibrators set 235454 Component: No

Z-1301-2026
Recall number
Z-1301-2026
Initiated
December 11, 2025
Classification
Class II
Status
Ongoing
Quantity
23 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low specificity when compared back to FDA-cleared CSF ratio/PET imaging, due to falsely elevated positive ratio or falsely elevated indeterminate ratio results causing an incorrect classification of a patient as having amyloid pathology associated with Alzheimer s disease (positive result) and/or requiring more clinical testing of that patient (indeterminate result).

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low specificity when compared back to FDA-cleared CSF ratio/PET imaging, due to falsely elevated positive ratio or falsely elevated indeterminate ratio results causing an incorrect classification of a patient as having amyloid pathology associated with Alzheimer s disease (positive result) and/or requiring more clinical testing of that patient (indeterminate result).

Code information

Model/Catalog Number: CL0073; UDI-DI: 00869164000086; Distribution for Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio ("Ratio") is related to the "Ratio" Instructions for Use (IFU). Therefore, each consignee has 1 "Ratio" in which they purchase the component kits for.

Distribution pattern

US Nationwide distribution in the states of Arizona, California, Florida, Indiana, Kansas, Kentucky, Maryland, Massachusetts, Minnesota, New Jersey, Pennsylvania, Texas, Washington.

device · product 2 of 7

Product Name: Lumipulse G pTau 217 Plasma Immunoreaction Cartridges Model/Catalog Number: 81550 Software Version: Not Applicable Product Description: Lumipulse G pTau 217 Plasma Immunoreaction Cartridges (IRC): Lumipulse G pTau 217 Plasma is an assay system, including a set of immunoassay reagents (Lumipulse G pTau 217 Plasma IRC), for the quantitative measurement of pTau 217 in K2EDTA plasma specimens based on chemiluminescent enzyme immunoassay (CLEIA) technology by a specific two-step immunoassay method on the LUMIPULSE G System. 3 ¿ 14 Tests: 3 racks, 14 IRCs per rack, 1 test per IRC. Each IRC contains (abbreviated contents): 1. Antibody-Coated Particle Solution: (Contains anti-phosphorylated Tau (217) monoclonal antibody (mouse)-coated particles). This solution contains gelatin and turns into gel at 15¿C or lower, 2. Enzyme-Labeled Antibody Solution: (Contains alkaline phosphatase (ALP)-labeled anti-Tau monoclonal antibodies (mouse) conjugate), 3. Assay Specific Solution: (Contains chemical stabilizers in 50 mM Tris buffer) Component: Yes. Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio

Z-1302-2026
Recall number
Z-1302-2026
Initiated
December 11, 2025
Classification
Class II
Status
Ongoing
Quantity
1569 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low specificity when compared back to FDA-cleared CSF ratio/PET imaging, due to falsely elevated positive ratio or falsely elevated indeterminate ratio results causing an incorrect classification of a patient as having amyloid pathology associated with Alzheimer s disease (positive result) and/or requiring more clinical testing of that patient (indeterminate result).

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low specificity when compared back to FDA-cleared CSF ratio/PET imaging, due to falsely elevated positive ratio or falsely elevated indeterminate ratio results causing an incorrect classification of a patient as having amyloid pathology associated with Alzheimer s disease (positive result) and/or requiring more clinical testing of that patient (indeterminate result).

Code information

Model/Catalog Number: 81550; UDI-DI: 05414470815500; Lot Numbers: D4C6028U, D4C6029U, D4C6051U, D4C6052U;

Distribution pattern

US Nationwide distribution in the states of Arizona, California, Florida, Indiana, Kansas, Kentucky, Maryland, Massachusetts, Minnesota, New Jersey, Pennsylvania, Texas, Washington.

device · product 3 of 7

Product Name: Lumipulse G pTau 217 Plasma Calibrators Model/Catalog Number: 81557 Software Version: Not Applicable Product Description: Lumipulse G pTau 217 Plasma Calibrators: The pTau 217 plasma concentration of a specimen is automatically calculated from the calibration curve, which is also automatically calculated from calibration data. The result of the calculation is reported in pg/mL. Lumipulse G pTau 217 Plasma Calibrators: Liquid (Frozen), 1x1.5 mL (5 Concentrations) CAL 1 0 pg/mL pTau 217 Plasma Calibrator CAL 2 0.250 pg/mL pTau 217 Plasma Calibrator CAL 3 1.000 pg/mL pTau 217 Plasma Calibrator CAL 4 5.000 pg/mL pTau 217 Plasma Calibrator CAL 5 10.000 pg/mL pTau 217 Plasma Calibrator Contains Tris buffer with protein (bovine) and chemical stabilizers. Preservative: 0.1% ProClin 300, 0.05% ProClin 950 Component: Yes. Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio

Z-1303-2026
Recall number
Z-1303-2026
Initiated
December 11, 2025
Classification
Class II
Status
Ongoing
Quantity
117 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low specificity when compared back to FDA-cleared CSF ratio/PET imaging, due to falsely elevated positive ratio or falsely elevated indeterminate ratio results causing an incorrect classification of a patient as having amyloid pathology associated with Alzheimer s disease (positive result) and/or requiring more clinical testing of that patient (indeterminate result).

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low specificity when compared back to FDA-cleared CSF ratio/PET imaging, due to falsely elevated positive ratio or falsely elevated indeterminate ratio results causing an incorrect classification of a patient as having amyloid pathology associated with Alzheimer s disease (positive result) and/or requiring more clinical testing of that patient (indeterminate result).

Code information

Model/Catalog Number: 81557; UDI-DI: 05414470815579; Lot Numbers: D5C6094U, D5C6125U;

Distribution pattern

US Nationwide distribution in the states of Arizona, California, Florida, Indiana, Kansas, Kentucky, Maryland, Massachusetts, Minnesota, New Jersey, Pennsylvania, Texas, Washington.

device · product 4 of 7

Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Product Description: Lumipulse pTau 217 Plasma Controls: This quality control product can be used as an objective judgement of the laboratory s procedures and personnel techniques when analyzing pTau 217 in K2EDTA plasma on the LUMIPULSE G System. It is a valuable tool to assess good laboratory practices. Two levels of control are available to compare observations with expected ranges therefore assuring consistent performance of the testing system within the clinical range. Lumipulse pTau 217 Plasma Controls: Liquid (Frozen), 2 ¿2 Concentrations L1 Level 1 (2¿ 1.5 mL) L2 Level 2 (2¿ 1.5 mL) Contains Tris buffer with protein (bovine) and chemical stabilizers. Preservative: 0.1% ProClin 300, 0.05% ProClin 950 Component: Yes. Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio

Z-1304-2026
Recall number
Z-1304-2026
Initiated
December 11, 2025
Classification
Class II
Status
Ongoing
Quantity
148 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low specificity when compared back to FDA-cleared CSF ratio/PET imaging, due to falsely elevated positive ratio or falsely elevated indeterminate ratio results causing an incorrect classification of a patient as having amyloid pathology associated with Alzheimer s disease (positive result) and/or requiring more clinical testing of that patient (indeterminate result).

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low specificity when compared back to FDA-cleared CSF ratio/PET imaging, due to falsely elevated positive ratio or falsely elevated indeterminate ratio results causing an incorrect classification of a patient as having amyloid pathology associated with Alzheimer s disease (positive result) and/or requiring more clinical testing of that patient (indeterminate result).

Code information

Model/Catalog Number: 81558; UDI-DI: 05414470815586; Lot Numbers: D6C6092U, D6C6092U1, D6C6123U;

Distribution pattern

US Nationwide distribution in the states of Arizona, California, Florida, Indiana, Kansas, Kentucky, Maryland, Massachusetts, Minnesota, New Jersey, Pennsylvania, Texas, Washington.

device · product 5 of 7

Product Name: Lumipulse G ¿-Amyloid 1-42-N Plasma Immunoreaction Cartridges Model/Catalog Number: 235447 Software Version: Not Applicable Product Description: Lumipulse G ¿-Amyloid 1-42-N Plasma Immunoreaction Cartridges (IRC): Lumipulse G ¿-Amyloid 1-42-N Plasma is an assay system, including a set of immunoassay reagents (Lumipulse G ¿-Amyloid 1-42-N Plasma IRC), for the quantitative measurement of ¿-Amyloid 1-42 in K2EDTA plasma specimens based on chemiluminescent enzyme immunoassay (CLEIA) technology13 by a specific two-step immunoassay method on the LUMIPULSE G System. Lumipulse G ¿-Amyloid 1-42-N Plasma Immunoreaction Cartridges (IRC): 3 ¿ 14 Tests: 3 racks, 14 IRCs per rack, 1 test per IRC. Each IRC contains (abbreviated contents): 1. Antibody-Coated Particle Solution: (Contains anti-¿-amyloid 1-42 monoclonal antibody (mouse)-coated particles). This solution contains gelatin and turns into gel at 15¿C or lower, 2. Enzyme-Labeled Antibody Solution: (Contains alkaline phosphatase (ALP)-labeled anti-¿-amyloid 1-42 monoclonal antibodies (mouse) conjugate), 3. Assay Specific Solution: (Contains detergents in 50 mM MES buffer) Component: Yes. Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio

Z-1305-2026
Recall number
Z-1305-2026
Initiated
December 11, 2025
Classification
Class II
Status
Ongoing
Quantity
1423 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low specificity when compared back to FDA-cleared CSF ratio/PET imaging, due to falsely elevated positive ratio or falsely elevated indeterminate ratio results causing an incorrect classification of a patient as having amyloid pathology associated with Alzheimer s disease (positive result) and/or requiring more clinical testing of that patient (indeterminate result).

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low specificity when compared back to FDA-cleared CSF ratio/PET imaging, due to falsely elevated positive ratio or falsely elevated indeterminate ratio results causing an incorrect classification of a patient as having amyloid pathology associated with Alzheimer s disease (positive result) and/or requiring more clinical testing of that patient (indeterminate result).

Code information

Model/Catalog Number: 235447; UDI-DI: 04987270235447; Lot Numbers: C5C6022U, C5C6045U;

Distribution pattern

US Nationwide distribution in the states of Arizona, California, Florida, Indiana, Kansas, Kentucky, Maryland, Massachusetts, Minnesota, New Jersey, Pennsylvania, Texas, Washington.

device · product 6 of 7

Product Name: Lumipulse G ¿-Amyloid 1-42-N Plasma Calibrators Model/Catalog Number: 235454 Software Version: Not Applicable Product Description: Lumipulse G ¿-Amyloid 1-42-N Plasma Calibrators: The ¿-Amyloid 1-42 plasma concentration of a specimen is automatically calculated from the calibration curve, which is automatically calculated from calibration data. The result of the calculation is reported in pg/mL. Lumipulse G ¿-Amyloid 1-42-N Plasma Calibrators: Liquid (Frozen), 1 ¿ 1.5 mL (4 Concentrations) CAL 1 0 pg/mL ¿-Amyloid 1-42-N Plasma Calibrator CAL 2 30 pg/mL ¿-Amyloid 1-42-N Plasma Calibrator CAL 3 100 pg/mL ¿-Amyloid 1-42-N Plasma Calibrator CAL 4 1000 pg/mL ¿-Amyloid 1-42-N Plasma Calibrator Contains Tris buffer with protein (bovine) and chemical stabilizers. Preservative: 0.1% ProClin 950. Component: Yes. Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio

Z-1306-2026
Recall number
Z-1306-2026
Initiated
December 11, 2025
Classification
Class II
Status
Ongoing
Quantity
128 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low specificity when compared back to FDA-cleared CSF ratio/PET imaging, due to falsely elevated positive ratio or falsely elevated indeterminate ratio results causing an incorrect classification of a patient as having amyloid pathology associated with Alzheimer s disease (positive result) and/or requiring more clinical testing of that patient (indeterminate result).

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low specificity when compared back to FDA-cleared CSF ratio/PET imaging, due to falsely elevated positive ratio or falsely elevated indeterminate ratio results causing an incorrect classification of a patient as having amyloid pathology associated with Alzheimer s disease (positive result) and/or requiring more clinical testing of that patient (indeterminate result).

Code information

Model/Catalog Number: 235454; UDI-DI: 04987270235454; Lot Numbers: D8C6022U, D8C6044U;

Distribution pattern

US Nationwide distribution in the states of Arizona, California, Florida, Indiana, Kansas, Kentucky, Maryland, Massachusetts, Minnesota, New Jersey, Pennsylvania, Texas, Washington.

device · product 7 of 7

Product Name: Lumipulse ¿-Amyloid Plasma Controls Model/Catalog Number: 81559 Software Version: Not Applicable Product Description: Lumipulse ¿-Amyloid Plasma Controls: This quality control product can be used as an objective judgement of the laboratory s procedures and personnel techniques when analyzing ¿-amyloid1-42 in K2EDTA plasma on the LUMIPULSE G System. It is a valuable tool to assess good laboratory practices. Two levels of control are available to compare observations with expected ranges therefore assuring consistent performance of the testing system within the clinical range. Lumipulse ¿-Amyloid Plasma Controls: Liquid (Frozen), 2 ¿2 Concentrations L1 Level 1 (2¿ 1.5 mL) L2 Level 2 (2¿ 1.5 mL) Contains Tris buffer with protein (bovine) and chemical stabilizers. Preservative: 0.1% ProClin 950 Component: Yes. Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio

Z-1307-2026
Recall number
Z-1307-2026
Initiated
December 11, 2025
Classification
Class II
Status
Ongoing
Quantity
179 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low specificity when compared back to FDA-cleared CSF ratio/PET imaging, due to falsely elevated positive ratio or falsely elevated indeterminate ratio results causing an incorrect classification of a patient as having amyloid pathology associated with Alzheimer s disease (positive result) and/or requiring more clinical testing of that patient (indeterminate result).

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low specificity when compared back to FDA-cleared CSF ratio/PET imaging, due to falsely elevated positive ratio or falsely elevated indeterminate ratio results causing an incorrect classification of a patient as having amyloid pathology associated with Alzheimer s disease (positive result) and/or requiring more clinical testing of that patient (indeterminate result).

Code information

Model/Catalog Number: 81559; UDI-DI: 05414470815593; Lot Number: T8B6116U;

Distribution pattern

US Nationwide distribution in the states of Arizona, California, Florida, Indiana, Kansas, Kentucky, Maryland, Massachusetts, Minnesota, New Jersey, Pennsylvania, Texas, Washington.