openFDA Device Enforcement
Report-date coverage
June 20, 2012–July 08, 2026
Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
openFDA Device Recall
event_date_initiated coverage
June 01, 1997–June 30, 2026
Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.
Product dates, classifications, firms, and source wording remain attached to their individual rows.
Opening this dossier never hides products that did not match a prior timeline filter.
It was identified that eh Universal Beam Triggering Interface (UBTI) is disabled in the Therapy Safety System (TSS) configuration. In this condition, the TSS does not interrupt the beam delivery if the Healthy Signal from an external gating device is lost. This may lead to a risk of mistreatment when external gating devices are used.
These labels are deterministic app interpretations, not FDA categories.
It was identified that eh Universal Beam Triggering Interface (UBTI) is disabled in the Therapy Safety System (TSS) configuration. In this condition, the TSS does not interrupt the beam delivery if the Healthy Signal from an external gating device is lost. This may lead to a risk of mistreatment when external gating devices are used.