Event 98411

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
February 02, 2026
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Howmedica Osteonics Corp.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

MICS3 Angled Sagittal Saw Attachment; Part Number: 210490

Z-1572-2026
Recall number
Z-1572-2026
Initiated
February 02, 2026
Classification
Class II
Status
Ongoing
Recalling firm
Howmedica Osteonics Corp.
Quantity
1098 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
A potential issue was identified with the torque strength on the external screws of the MICS3 Angled Sagittal Saw Attachment. These screws, which secure the cover to the Angled Saw Housing, may loosen during use and detach from the attachment.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

A potential issue was identified with the torque strength on the external screws of the MICS3 Angled Sagittal Saw Attachment. These screws, which secure the cover to the Angled Saw Housing, may loosen during use and detach from the attachment.

Code information

Part Number: 210490; UDI: (01)07613327523874(10)TU6563205RP; Lot numbers: ALL;

Distribution pattern

Domestic: Nationwide Distribution; International: Australia, Germany, Belgium, France, Spain UK, New Zealand;