Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 98461

2 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
February 25, 2026
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Synthes (USA) Products LLC

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

2 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 2

2.4 VOLT TAP LOCKING SCREW TAP TO 90MM. Part Number: 03.424.124-US.

Z-1702-2026
Recall number
Z-1702-2026
Initiated
February 25, 2026
Classification
Class II
Status
Ongoing
Quantity
77 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
A full lot swap between VOLT Locking Screw Tap 03.424.124 lot 68900P9 and Cortical Screw Tap 03.424.126 lot 68982P8, resulting in the bones taps having an incorrect thread.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process change control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

A full lot swap between VOLT Locking Screw Tap 03.424.124 lot 68900P9 and Cortical Screw Tap 03.424.126 lot 68982P8, resulting in the bones taps having an incorrect thread.

Code information

Part Number: 03.424.124-US. GTIN: 10886982335944. Lot Number: 68900P9.

Distribution pattern

US Nationwide distribution in the states of AR, AZ, CA, CO, FL, IN, LA, MD, MN, NC, NJ, OH, OK, OR, PA, TN, TX, UT, VA, WA, WI.

device · product 2 of 2

2.4 VOLT(TM) TAP CORTEX SCREW TO 90MM. Part Number: 03.424.126-US.

Z-1703-2026
Recall number
Z-1703-2026
Initiated
February 25, 2026
Classification
Class II
Status
Ongoing
Quantity
33 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
A full lot swap between VOLT Locking Screw Tap 03.424.124 lot 68900P9 and Cortical Screw Tap 03.424.126 lot 68982P8, resulting in the bones taps having an incorrect thread.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process change control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

A full lot swap between VOLT Locking Screw Tap 03.424.124 lot 68900P9 and Cortical Screw Tap 03.424.126 lot 68982P8, resulting in the bones taps having an incorrect thread.

Code information

Part Number: 03.424.126-US. GTIN: 10886982335951. Lot Number: 68982P8.

Distribution pattern

US Nationwide distribution in the states of AR, AZ, CA, CO, FL, IN, LA, MD, MN, NC, NJ, OH, OK, OR, PA, TN, TX, UT, VA, WA, WI.