device · product 1 of 7
Centurion medical convenience kits, containing Webcol Large Alcohol Prep Pads, labeled as: Universal Central/Art Line Insertion Kit, Medline Kit SKU ART155A
- Recall number
- Z-2039-2026
- Initiated
- March 16, 2026
- Classification
- Class II
- Status
- Ongoing
- Recalling firm
- Medline Industries, LP
- Quantity
- 120 kits
App-derived interpretation
Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing Cardinal Health Webcol Large Alcohol Prep Pads. Cardinal Health has deemed Webcol Large Alcohol Prep Pads to be non-sterile following the discovery of a contaminant (Paenibacillus phoenicis) during a routine sterilization dose audit.
Official device-enrichment evidence · Unknown
An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.
Inspect official wording and provenance
Reason for recall
Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing Cardinal Health Webcol Large Alcohol Prep Pads. Cardinal Health has deemed Webcol Large Alcohol Prep Pads to be non-sterile following the discovery of a contaminant (Paenibacillus phoenicis) during a routine sterilization dose audit.
Code information
UDI/DI each 10653160316459, UDI/DI case 00653160316452, Lot Numbers: 2025100890, 2026011290.
Distribution pattern
US Nationwide distribution.