Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 98894

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
May 06, 2026
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Stryker Corporation

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Neptune Smart Docking Station (120V), Catalog Number 0770-100-000, part of the Neptune Waste Management System

Z-2282-2026
Recall number
Z-2282-2026
Initiated
May 06, 2026
Classification
Class II
Status
Ongoing
Recalling firm
Stryker Corporation
Quantity
1 unit

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
It has been identified that the impacted docking station (SN 2521008873) was not product released from manufacturing and may not have completed required testing prior to being shipped to your facility. There is a potential risk of electrical shock due to the device s safety testing being unverifiable.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Nonconforming Material/Component

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

It has been identified that the impacted docking station (SN 2521008873) was not product released from manufacturing and may not have completed required testing prior to being shipped to your facility. There is a potential risk of electrical shock due to the device s safety testing being unverifiable.

Code information

GTIN 07613327634914, Serial Number 2521008873

Distribution pattern

US Nationwide distribution in the state of Indiana.