Ivenix Infusion System (IIS), Large Volume Pump, Model Number: LVP-0004
- Recall number
- Z-2224-2026
- Initiated
- May 06, 2026
- Classification
- Class I
- Status
- Ongoing
- Recalling firm
- Fresenius Kabi USA, LLC
- Quantity
- 18,444
App-derived interpretation
Fresenius Kabi has been made aware via customer complaints that dropping or severe jarring of the Ivenix LVP may cause damage to internal components that may not be apparent or readily visible and the devices therefore remained in service. More specifically, drops or jarring on a hard surface can cause the touchscreen input cable to loosen, resulting in unintended touchscreen behavior on the LVP, including random screen touches and/or lack of screen responsiveness to user input. Under the reported failure condition, unintended touch inputs may register at arbitrary screen locations without user interaction. These inputs occur intermittently and without a predictable pattern.
Official device-enrichment evidence · Sourced
Device Design
Inspect official wording and provenance
Reason for recall
Fresenius Kabi has been made aware via customer complaints that dropping or severe jarring of the Ivenix LVP may cause damage to internal components that may not be apparent or readily visible and the devices therefore remained in service. More specifically, drops or jarring on a hard surface can cause the touchscreen input cable to loosen, resulting in unintended touchscreen behavior on the LVP, including random screen touches and/or lack of screen responsiveness to user input. Under the reported failure condition, unintended touch inputs may register at arbitrary screen locations without user interaction. These inputs occur intermittently and without a predictable pattern.
Code information
Model Number: LVP-0004; UDI-DI: 00811505030320; All Serial/Lot Numbers:
Distribution pattern
Domestic: CA, CO, FL, GA, ID, IL, MD, MI, MI , MN, MO, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI, WI;