Skip to content
Recall Observatory FDA recall evidence

Device product

RayCare 2024A SP4. Oncology Information Systems.

Z-2647-2026

May 20, 2026

Class II

Product summary

Firm
RAYSEARCH LABORATORIES AB
Event
Event 99141
Status
Ongoing
Classification
Class II
Quantity
3 units
Official record key
device-enforcement:Z-2647-2026

Official wording

Reason: Beam set delivery note issues in which there is potential data loss if the beam set delivery note is modified simultaneously and/or if a note is updated during a treatment session, the user is not notified of the change.

Code information: Software Version 8.0.4.60077, UDI 0735000201089120241218, Expiration Date: 2030-01-24.

Distribution pattern: Worldwide - US Nationwide distribution.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Beam set delivery note issues in which there is potential data loss if the beam set delivery note is modified simultaneously and/or if a note is updated during a treatment session, the user is not notified of the change.