INZONE DETACHMENT SYSTEM, REF: M00345100950

Z-2552-2026

April 06, 2026

Class II

Product summary

Firm
Stryker Neurovascular
Event
Event 98630
Status
Ongoing
Classification
Class II
Quantity
44,937
Official record key
device-enforcement:Z-2552-2026

Official wording

Reason: Embolization coil detachment system devices may experience premature battery drain causing devices to: 1) Not power on; 2) Power on with faint audible and visual indicators; 3) Be unable to detach a coil as intended, which may necessitate medical intervention if the procedure must be completed with an alternative technique.

Code information: UDI-DI: 04546540697950. Lots: WMP133638, WMP133871, WMP133872, WMP133988, WMP133989, WMP134016, WMP134132, WMP134133, WMP134234, WMP134235, WMP134255, WMP134256, WMP134380, WMP134381, WMP134432, WMP134433, WMP134518, WMP134519, WMP134609, WMP134610, WMP134794, WMP134795, WMP134939, WMP134940, WMP135187, WMP135243, WMP135302, WMP135314, WMP135354, WMP135355, WMP135387, WMP135417, WMP135473, WMP135475, WMP135491, WMP135492, WMP135623, WMP135720, WMP135727, WMP135807, WMP135865, WMP135897, WMP136084, WMP136107, WMP136190, WMP136200, WMP136611, WMP136642, WMP136736, WMP136748, WMP136852, WMP136894, WMP136985

Distribution pattern: Worldwide - US Nationwide distribution including in the states of PA, AZ, FL, CA, IL, AK, NY, MI, NJ, WI, SD, KY, AR, TX, AL, TN, MO, MA, IN, NE, WV, MT, CT, NC, VA, NV, CO, DC, DE, OH, ME, GA, WA, SC, GU, MN, OK, NH, MD, ND, UT, HI, OR, LA, MS, KS, NM, ID, IA, VT and the countries of ARGENTINA, AUSTRIA, BELGIUM, BRASIL, BULGARIA, CANADA, CANARY ISLANDS, CHILE, COLOMBIA, CROATIA, CZECH REPUBLIC, DENMARK, ESTONIA, FINLAND, FRANCE, GERMANY, GREECE, HONG KONG, HUNGARY, ICELAND, IRAN, IRELAND, ISRAEL, ITALIA, JAPAN, KOREA, LATVIA, LUXEMBOURG, MALAYSIA, MEXICO, NETHERLANDS, NORTHERN IRELAND, NORWAY, PHILIPPINES, POLAND, PORTUGAL, QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SOUTH PACIFIC, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Embolization coil detachment system devices may experience premature battery drain causing devices to: 1) Not power on; 2) Power on with faint audible and visual indicators; 3) Be unable to detach a coil as intended, which may necessitate medical intervention if the procedure must be completed with an alternative technique.